Puridest Mono Distillation Apparatus

09/2020

3

Before Initiation

Important information are marked in bold letters in these instructions, safety indications are additionally marked by the
following warning symbols and mandatory signs.

  Read and observe the
  operating instructions

Warning of hot liquids

and vapour

Warning of hot surfaces

Warning of dangerous
electrical voltage

General warning

Before maintenance and repair

disconnect the unit all-pole from

the electrical mains (pull the plug
from the socket).

4

Set-up and Location of the Distillation Apparatus

All  free  openings  of the  water  connections on  the  right-hand  side  of the  unit  are  taped  close. These
transport  protections  have  to  be  removed  before  installation. Place  on  solid,  even  and  level  surfaces
inside  buildings  only.  Make  sure  to  place  the  unit  only  on  a  watertight,  temperature-resistant,  non-
flammable surface. The location must provide sufficient space as well as the necessary carrying capacity
for the unit (unit weight as per technical data, chapter 11 of this manual, plus weight of the filling. The
unit is not suitable for use in potentially explosive areas.

5

Operating Voltage

The Distillation Apparatus must be connected to the mains through a correctly installed shock-proof
socket or through an on-site main switch. The Distillation Apparatus is a protection class I electrical
appliance, a connection to the earth conductor (PE) must be ensured. For information on the required
mains fuse please view Technical Data 11 of this manual.

The electrical connection must ensure an all-pole separation of the Distillation Apparatus from the
mains at any time. The mains connection cable must not touch any hot surfaces of the unit. It may not
lead underneath the unit. The main switch of the unit must be off (position O). The voltage on the
nameplate (at the back of the unit) must be identical to the mains voltage. If they are identical, connect
the unit to the mains. For further information, please also refer to chapter 13 of these operating in-
structions "Connection to the Mains Supply".

Summary of Contents for PD 2

Page 1: ...Operating Instructions Puridest Mono Distillation Apparatus PD 2 PD 4 ...

Page 2: ...2 22 Puridest Mono Distillation Apparatus 09 2020 ...

Page 3: ...ity and composition of the feed water Before installation please check whether contents of package are in good order and complete Should you note any damages or have any reasons for complaint please contact your supplier or directly LAUDA GFL Gesellschaft für Labortechnik mbH Schulze Delitzsch Str 4 5 30938 Burgwedel Germany Telefon 49 0 5139 9958 0 Fax 49 0 5139 9958 21 E Mail info lauda gfl de I...

Page 4: ...4 22 Puridest Mono Distillation Apparatus 09 2020 ...

Page 5: ...ng Water Temperature 10 Control Thermometer ill C 10 7 1 Water inlet and Setting the Cooling Water Temperature ill D and E 10 7 2 Degassing and Distillate Withdrawal ill F 10 7 3 8 Functional Description 11 9 Maintenance Service and Removal of Malfunctions 11 Descaling 11 9 1 Cleaning 12 9 2 Re Initiation after Low Water 12 9 3 10 Disposal of Old Units 12 11 Technical Data 13 12 Circuit diagram 14...

Page 6: ...6 22 Puridest Mono Distillation Apparatus 09 2020 ...

Page 7: ...ect operation of the Distillation Apparatus can be guaranteed The units may only be installed and operated by persons who have made themselves familiar with these operating instructions Caution Both the outer housing of the boiler and the steam condenser get strongly heated up during operation Both parts may only be touched after they have cooled down or when using suitable safety gloves Improper ...

Page 8: ...as well as the necessary carrying capacity for the unit unit weight as per technical data chapter 11 of this manual plus weight of the filling The unit is not suitable for use in potentially explosive areas 5 Operating Voltage The Distillation Apparatus must be connected to the mains through a correctly installed shock proof socket or through an on site main switch The Distillation Apparatus is a ...

Page 9: ...ct a temperature resistant hose with a maximum length of 1 5 m to the cooling water outlet The hose must lead into a drain on a lower level and must have a slope on its complete length Make sure the cooling water can drain with out back draughts Caution The cooling water leaves the Distillation Apparatus with a temperature up to 70 C Danger of scaldings Destillate Outlet 6 3 Use a temperature resi...

Page 10: ...op valve of the mains water supply a little Check the power supply and switch on the main switch The pilot lamp in the main switch lights up As soon as the water in the boiler reaches boiling point regulate the cooling water supply on the tap of the mains water supply so as to set the temperature of the draining cooling water to approx 60 C to 70 C The tem perature can be checked with the inserted...

Page 11: ...Distillation Apparatus as well as before cleaning procedures discon nect the unit from the mains Unplug the unit or switch off the main switch Danger of electrical shock Descaling 9 1 Depending on the hardness of the tap water the Distillation Apparatus must be cleaned from scale formations in regular intervals To do so close the on site water supply valve and lift the condenser from the boiler A ...

Page 12: ... with a pen until a clicking sound can be heard The LAUDA Puridest Distillation Apparatus is produced with first class materials Nevertheless the unit should only be subjected to mechanical strain in sensible limits Make sure that no fluids come into contact with cable connections or the electrical parts inside the unit The powder coated surfaces of the unit can be cleaned with mild detergents if ...

Page 13: ...with capillary tube sensor Switch off temperature 135 C 15 K Required cooling water pressure 1 2 bar 14 5 psi 29 0 psi 1 2 bar 14 5 psi 29 0 psi Electrical connection Mains connection 230 V 10 50 60 Hz 2 0 kW Schuko plug CEE 7 7 230 V 10 50 60 Hz 3 0 kW Schuko plug CEE 7 7 Caution Mains voltage deviations even within the mentioned limits influence the quantity of distillate produced Mains fuse On ...

Page 14: ...ircuit diagram E1 Tubular heating element F1 Fuse 10 Amp inert PD 2 Fuse 15 Amp inert PD 4 F2 Fuse 10 Amp inert PD 2 Fuse 16 Amp inert PD 4 F3 Low water cut off thermostat K1 Contactor PD 4 S1 Main switch Circuit diagram PD 2 12 1 Circuit diagram PD 4 12 2 ...

Page 15: ...ctive conductor terminal Colour coding of the mains cable Mains supply ge gr yellow green PE Protective earth bl blue N sw black L1 Models PD 2 and PD 4 supplied for 230 V see information on the nameplate can be connected to all power supplies of 220 V or 230 V Electrical fuses 13 1 Model Power Power consumption at mains voltage Mains fuse F4 F5 PD 2 2 0 kW 8 7 A at 230 V 10 Amp PD 4 3 0 kW 13 0 A...

Page 16: ... unit F4 Mains fuse on site F5 Mains fuse on site S2 Main switch on site PD 2 und PD 4 with mains supply 230 V N PE 50 60 Hz connected through 3 pole shock proof plug system PD 2 und PD 4 for 230 V with mains supply 400 V 3 N PE 50 60 Hz connected through on site main switch PD 2 und PD 4 with mains supply 220 V 3 PE 50 60 Hz connected through 3 pole shock proof plug system ...

Page 17: ...0127 Main switch 4 A000091 Contactor 5 A000116 Stand 6 A000092 Low water cut off thermostat 7 0030097 Mains connection cable 7 1 A000125 Fuse holder 2 pole 7 2 0012093 Mains fuse 10 Amp inert PD 2 0012092 Mains fuse 15 Amp inert PD 4 8 0013104 Thermometer 9 A000115 Hose tap 7 1 7 2 Please always state model and serial no of the Distillation Apparatus when placing an order for spare parts informati...

Page 18: ...pre cleaning the tap water and to protect the unit from premature contamination Complete with connections for pressure hose inch including filter candle The candle should be replaced after maximum 6 months Pre Filter with candle Spare candle Part No A000133 Part No A000134 Wall bracket for one filter or for filter combinations of two or three filters including sleeves to connect the filters as wel...

Page 19: ...09 2020 Puridest Mono Distillation Apparatus 19 22 ...

Page 20: ...20 22 Puridest Mono Distillation Apparatus 09 2020 ...

Page 21: ...09 2020 Puridest Mono Distillation Apparatus 21 22 ...

Page 22: ...LAUDA GFL Gesellschaft für Labortechnik mbH Schulze Delitzsch Straße 4 5 30938 Burgwedel Deutschland Tel 49 0 5139 9958 0 Fax 49 0 5139 9958 21 E Mail info lauda gfl de Internet www GFL de ...

Reviews: