09/2020
Puridest Mono Distillation Apparatus
15 / 22
13
Connection to the mains supply
Distillation Apparatus models PD 2 and PD 4 are supplied with a pre-assembled, cast-on shock-proof plug (CEE 7/7).
Make sure to connect to a protective conductor terminal.
Colour coding of the mains cable
Mains supply
ge/gr – yellow/green
PE (Protective earth)
bl – blue
N
sw – black
L1
Models PD 2 and PD 4 supplied for 230 V (see information on the nameplate) can be connected to all power supplies of
220 V or 230 V.
Electrical fuses
13.1
Model
Power
Power consumption at mains voltage *
Mains fuse (F4, F5)
PD 2
2.0 kW
8.7 A at 230 V
10 Amp
PD 4
3.0 kW
13.0 A at 230 V
16 Amp