6.7.3
Write commands
The write commands are data specifications to the thermostat.
Command
Meaning
OUT_SP_00_XXX.XX
Setpoint transfer with max. 3 digits before the dec-
imal point and max. 2 digits afterwards
OUT_SP_04_XXX
[Hi] Upper limit of flow temperature
OUT_SP_05_XXX
[Lo] Lower limit of flow temperature
START
Switches on device (from standby)
STOP
Switches device to standby (pump, refrigeration unit
off).
Note the following:
For "_", " " (space character) is also permitted.
Response from the thermostat “OK” or “ERR_X” in the case of
an error.
-XXX.XX
-XXX.X
-XXX.
-XXX
XXX.XX
XXX.X
XXX.
XXX
-XX.XX
-XX.X
-XX.
-XX
XX.XX
XX.X
XX.
XX
-X.XX
-X.X
-X.
-X
X.XX
X.X
X.
X
-.XX
-.X
.XX
.X
6.7.4
Read commands
The following read commands are data requests to the thermostat.
Command
Meaning
IN_PV_00
Query of the bath temperature (outlet temperature)
IN_SP_00
Query of the temperature setpoint
IN_SP_04
Query of the outlet temperature limit Hi
IN_SP_05
Query of the outlet temperature limit Lo
TYPE
Query of the device type (response = "MC").
VERSION
Query of the software version number
Permitted data formats
Operation
Microcool