Starting up
34
−
Turn the potentiometer slowly anticlockwise.
!
The unit
must switch off at the bath temperature.
−
Step 1 - 2 (see above) must follow.
−
Set the overtemperature switch-off point again above the
bath temperature and wait until the indication TEMP
appears on the display, then
−
reset with the key.
6.5.2
Low-level protection and testing
$
$
$
$
−
Double signal tone sounds, if the bath liquid falls so much
that the heater is no longer covered with liquid completely.
1. The display shows LEVEL (low-level) and the fault triangle
is flashing
!
heating is switched off on both poles,
!
pump is switched off by the electronics
2. Top up the bath
➾
Section 5.2 and reset with the key.
☞
−
If necessary repeat several times in case that the pump does not purge
immediately.
−
Testing at regular intervals by lowering the bath level. Place a hose on the pump
connector and pump some of the bath liquid into a suitable container.
−
Step 1 - 2 must follow.
−
Bath temperature during this test not below 0°C or higher than 50°C, otherwise
danger of burn injuries !
−
If there is any irregularity when testing the safety devices, switch off the unit
immediately and pull out the mains plug !
−
Have the unit checked by the LAUDA service or the local service organisation!