Aqualine
20
Maintenance
15.03.18/ YACE0086
Water must not penetrate the control unit!
Carry out appropriate detoxification if dangerous material has been spilled on or in-
side the unit.
Method of cleaning and detoxification are decided by the special knowledge of the
user. In case of doubt please contact the manufacturer.
8.4 Maintenance and repair
Withdraw the equipment mains plug before all service and repair work!
Repair on the control unit must only be carried out by properly qualified personnel!
LAUDA Aqualine water baths are maintenance-free. If the heat carrier liquid becomes dirty it has to be
replaced (
If a fuse blows (
supply indication not alight) fit only fuses as specified.
(2 off T (slow blow) 12.5 A (EES 014); 1 off T (slow blow) 0.2 A (EES 069); Size 6.3 x
32
fuses located in the unit).
Changing the fuses:
Before opening, withdraw the mains plug, pick up the gable cover and empty the bath.
Remove the rubber buffer out of the four feet. Release the screws in the feet and in the bottom
panel with a cross-head screwdriver, loosen the earthing cable and remove the bottom panel.
The fuses are located behind the heat sink at the rear of the unit.
Replace the blown fuse and reassemble the water bath in the reverse sequence.
Printed circuit
board
UL 581
F3; T 0.2 A
(EES 069)
F1; F2,
T12.5 A
(EES 014)
Before sending us the unit, please contact our technical service. (
If the equipment has to be returned to the factory, please ensure that it is carefully and
properly packed. LAUDA accepts no responsibility for damage due to unsatisfactory
packing.