MultiMeter-Compact
25
Subject to technical alterations. 19W09
EU directives and disposal
This device complies with all necessary standards
for the free movement of goods within the EU.
This product is an electric device and must
be collected separately for disposal according
to the European Directive on waste electrical
and electronic equipment.
Further safety and supplementary notices at:
http://laserliner.com/info?an=ADV
Frequency
9.999 Hz
± (1.5% rdg ± 5 digits)
99.99 Hz
999.9 Hz
± (1.2% rdg ± 3 digits)
9.999 kHz
99.99 kHz
999.9 kHz
9.999 MHz
± (1.5% rdg ± 4 digits)
Duty factor
0.1%…99.9%
± (1,2% rdg ± 2 digits)
Diode test
0.3 mA
± (10% rdg ± 5 digits)
Polarity
Sign for negative polarity
LCD
0 … 3999
Fuse
10 A/600 V quick-acting,
240 A2/s (6.35 x 31.8 mm)
Protection class
II, double insulation
Overvoltage
CAT III - 1000 V, CAT IV - 600 V
Pollution degree
2
Test standards
EN 61326, EN 61010-1, EN 61010-2-031
Operating conditions
0°C ... 55°C, max. humidity 80% rH,
no condensation, max. working altitude
2000 m above sea level
Storage conditions
0°C ... 60°C,
max. humidity 80% rH
Power supply
1 x 9 V battery (NEDA 1604, IEC 6F22)
Dimensions (W x H x D)
150 mm x 70 mm x 48 mm
Weight
255 g (incl. battery)
EN