MasterPlane-Laser 3G
19
A1
A2
2.
1.
A3
A2
A2
A1
4.
3.
A3
A2
<
0,2 mm / m = OK
Performing the calibration check
3.
Position the device as near as possible to the wall at the height
of point A1.
4.
Turn the device through 180° and mark point A3.
The difference between points A2 and A3 is the tolerance.
1.
Mark point A1 on the wall.
2.
Turn the device through 180° and mark point A2.
You now have a horizontal reference between points A1 and A2.
Preparing the calibration check
It is possible for you to check the calibration of the laser. To do this, position
the device
midway
between 2 walls, which must be at least 5 m apart.
Switch the device on
(Laser cross ON)
. The best calibration results are
achieved if the device is mounted on a tripod.
Due to the special optics required to generate a continuous 360° laser
line, the underlying technology may cause differences in brightness in
different areas of the line. This may lead to different ranges in hand
receiver mode.
!
EN