Error Codes
An error code will be displayed if the RD 1000 detects a problem with a
measurement. Depending upon the current function, error codes are displayed in
either the in-scope LED or the external LCD display. The table below lists and
explains the possible error codes.
Code
Explanation
Remedy
E01
Unstable inclination value. The
instrument is too unsteady to
produce an accurate reading.
•
Steady the instrument
before releasing the
TRIGGER button.
E02
Calculation error. There was an
error in performing an internal
calculation caused by improper
measurement geometry. Likely
caused by a incorrectly entered
data value, or an incorrect target
shot from an external laser
device.
•
Re-enter or re-shoot the
data.
E03
Data communication error. There
was an error in a received serial
data string from an external
device.
•
Verify that the laser and the
RD 1000 are using the same
units.
•
Verify the laser mode
(it should be in HD or VD)
and re-shoot.
E04
System memory error. There was
some type of failure of the
internal system memory. This
represents a memory checksum
failure of factory stored
parameters.
•
If the error persists, contact
Laser Technology, Inc.
E05
Data entry error. An improper
value was entered during a
manual data edit operation (the
entered data was outside of
acceptable value limits).
•
Check value and re-enter
data.
E06
Unable to display the input value
in the in-scope LED or external
LCD.
•
Re-enter data.
•
Hold the instrument within
tilt limits.
Unable to display the calculated
value.
•
Press the ENTER button to
continue with the
calculation.