9/29/09
Using ID Setting Files
7-7
NOTE: If Gain is set to -10 or -20 dB
and the microphone type is not
condenser then the gain is set to 0 dB
and the warning “Setting Conflict”
is displayed.
-20, -10, 0, +10, +20, +30, +40, +50
Use the
u
and
d
keys to highlight the desired value
of gain and press
c
Check key.
Range Setting - (LOG, ISM)
The Range setting is only available
to the LOG and ISM instrument.
Range
selects the measurement range you desire for the
instrument. This setting provides the following three
options:
NOTE: If Range is set to “Low” and
the microphone type is not condenser
then the range is set to “Normal”
and the warning “Setting Conflict”
is displayed.
Normal -
The instrument’s available dynamic
range is positioned in the “normal” range which is
typically 20dB to 130dB when using a Lar-
son•Davis high-sensitivity microphone.
Low -
The instrument’s input circuitry provides an
additional 20dB of gain which has the effect of
moving the instrument’s available dynamic range
downward relative to the “normal” range. The net
effect is that you can measure lower-level sounds
effectively. The degree of benefit varies depending
upon the weighting or filter that you are using.
High Range is not available for
electret or direct transducer types
(see “Transducer Setting” below).
High -
The instrument’s sensitivity is reduced by
20dB, which has the effect of moving the instru-
ment’s available dynamic range upward relative to
the “normal” range. Therefore you can measure
higher-level sounds effectively. The benefit varies
depending upon the weighting or octave filter that
you are using.
Summary of Contents for System 824
Page 1: ...Model 824 Precision Sound Level Meter Real Time Analyzer Reference Manual...
Page 2: ......
Page 3: ...I824 01 Rev P Technical Reference Manual Larson Davis Model 824...
Page 58: ...2 18 824 Reference Manual 9 29 09...
Page 120: ...4 24 824 Reference Manual 9 29 09...
Page 194: ...5 74 824 Reference Manual 9 29 09...
Page 284: ...7 80 824 Reference Manual 9 29 09...
Page 312: ...8 28 824 Reference Manual 9 29 09...
Page 330: ...9 18 824 Reference Manual 9 29 09...
Page 366: ...10 36 824 Reference Manual 9 29 09...
Page 460: ...A 4 824 Reference Manual 9 29 09 shown below Both versions work properly with the Model 824...
Page 538: ...A 82 824 Reference Manual 9 29 09...
Page 546: ...B 8 824 Reference Manual 9 29 09...
Page 569: ...9 29 09 C 23 Windscreen Response with Respect to No Windscreen...
Page 596: ...C 50 824 Reference Manual 9 29 09...
Page 630: ...E 10 824 Reference Manual 9 29 09...
Page 638: ...F 8 824 Reference Manual 9 29 09...
Page 648: ...G 10 824 Reference Manual 9 29 09...
Page 672: ...iv xxiv Index Model 831 Technical Reference Manual...