SoundAdvisor Model 831C
Creating a Custom Setup File On the Meter
6-7
Step 13
If the weather option is enabled on your meter, use the right (
//*
) softkey
to navigate to the
Weather
FIGURE 6-10
Setup Manager: (Optional) Weather Tab
1.
This section applies only when weather accessories are
connected to the 831C. On the Setup Manager Weather tab,
set the
Weather
station type and any other applicable values
for the measurement of wind speed, wind direction, tempera
-
ture and humidity.
2.
If the wind speed is above the
Threshold
, it is considered
windy.
3.
If
Excd. Pause
is set to
Yes
, when the wind speed exceeds
Excd. Level
, sound exceedances are held off.
4.
During a wind exceedance (with sound exceedance
paused), the wind speed must drop below the
Hysteresis
before sound exceedance pause is turned off.
6.2.3 Saving the Setup File
Before you begin:
•
Complete
6.2.1 Creating a New Setup File On the Meter
•
Complete
6.2.2 Customizing Setup Values On the Setup Manager
Step 1
To save your Setup and exit the Setup Manager, select
Close
(
/*/
).
Step 2
In the Save dialog, choose
Yes
to save your settings.
Step 3
Highlight and press
5
on the name of the Setup you created.
Step 4
In the pop-up menu that appears,
Set to Active
is highlighted.
Press
5
to
select that option. The next measurement runs using the Setup you just
created.
TAKE NOTE
You can also complete the setup process in G4 LD Utility, which provides a
simple, file transfer-style interface for configuring, saving, and transferring
Setup files to your PC or to another meter. For more information, refer to the
G4 LD Utility Reference Manual
.
6.2.4 Choosing a Run Mode
Before you begin:
•
Complete steps 1–7 of section
6.2 Creating a Custom Setup File On
.
1.
2.
3.
4.