Model 831 Manual
Making a Measurement
15-5
Measurement State Sequence
This section presents the sequence of Model 831 screen
displays and actions taking place during an a measurement.
FIGURE 15-5 RT-60 Status Screen
Initiate Measurement
Initiate a measurement by pressing the
9
(RUN/PAUSE)
key, at which time the icon shown in FIGURE 15-6 will
appear in the lower left corner of the screen. This lasts as
long as the configured Exit Time. No data is taken while this
icon is shown.
FIGURE 15-6 Exit Prompt
Background Noise Measurement
After the Exit Time has passed, the background noise is
measured, as indicated by the icon shown in FIGURE 15-7.
The background noise level is measured for 5 seconds and is
then used for determining signal-to-noise ratio on successive
decays. The background noise level is measured for every
position, i.e., every time "Run" is pressed.
Trigger Level
Trigger Level - 45 dB
Summary of Contents for 831
Page 1: ...Model 831 Sound Level Meter Manual...
Page 2: ...I831 01 Rev O Manual Larson Davis Model 831 Manual...
Page 123: ...5 44 Any Level Display Model 831 Manual...
Page 135: ...6 12 Run Modes Without Measurement History Model 831 Manual...
Page 155: ...7 20 Recovery After Improper Shutdown Model 831 Manual...
Page 209: ...11 22 Markers Model 831 Manual...
Page 221: ...12 12 Link to Time History Model 831 Manual FIGURE 12 11 Time History Display...
Page 321: ...16 24 Sound Recording Playback Model 831 Manual...
Page 358: ...Model 831 Manual Displays 18 27 FIGURE 18 42 Measurement Displays FIGURE 18 43 Events Displays...
Page 403: ...20 28 Troubleshooting Model 831 Manual...
Page 413: ...21 10 Calibration When The Model 831 Is Locked Model 831 Manual...
Page 421: ...23 4 File System Model 831 Manual...
Page 501: ...A 64 Vibration Sensitivity Model 831 Manual...
Page 513: ...B 12 Sections 5 6 7 and 9 except 9 3 Model 831 Manual...
Page 525: ...C 12 Community Noise Descriptors Model 831 Manual...