![lanry TF1100-CH Operation & Maintenance Manual Download Page 39](http://html1.mh-extra.com/html/lanry/tf1100-ch/tf1100-ch_operation-and-maintenance-manual_3586861039.webp)
TF1100-CH transit-time handheld flowmeter manual
36
APPENDIX 2 DATA LOGGER USEGE
Meter Setting Before Using Data Logger
Before using the data logger, please check the following items of meter settings (otherwise,
the data logger will not work normally):
I. In Menu 60, it is time setup for meter, the format should be:
YY-MM-DD (For example, it is Aug 14, 2015 today, the
Menu 60 setting should be 15-08-14)
HH-MM-SS (Hour-minute-second, in 24 hour format)
II. In Menu 50
Select
Logger Options
ON
III. Menu51, you setup start time (later than Menu 60 time, had better at least later 2
minutes), or setting up start time as: Start **:**:** is also Okay.
or
IV. Menu 52,
Select
0. To RS-232
V. Menu 62
Select
Baudrate: =9600
Parity: None
Then insert the Data Logger module. The meter will restart, which means the data logger
works well.
IMPORTANT:
Restart the meter is a must before data logger working. Only insert the SD
card the meter will not restart, and the data logger will
stop recording data.
5.6.3 Read Data
When downloading data, it will create a new .csv file named for date,, the naming rules of the
M60
Date 15-08-14
Time 15:16:46
M50
Logger Options
ON
M51 Logger Setup
Start 15:20:00
Intervl 00:00:05
Go On **:**:**
M51 Logger Setup
Start **:**:**
Intervl 00:00:05
Go On **:**:**
M62
RS-232C Setup
Baudrate: =9600
Parity: None
M52
Data Output
0. To RS-232