TF1100-CH transit-time handheld flowmeter manual
33
APPENDIX 1 TRANSDUCER
GUIDE
RAIL
INSTALLATION
Introduction
The guide rail is designed for clamp-on transducer installation, it make the installation
simple, allocation precise, fixation stable, and it also can meet to multiple transducer
mounting metheds.
Preparation
1.
Before you attach the transducers you should first ensure that the proposed location
satisfies the distance requirements as
Table 2.1
, otherwise the resulting accuracy of the flow
readings may be affected.
2.
Prepare the pipe by degreasing it and removing any loose material or flaking paint in
order to obtain the best possible surface. A smooth contact between pipe surface and the
face of the transducers is an important factor in achieving a good ultrasound signal strength
and therefore maximum accuracy.
3. Parts prepare
Chins (2pcs)
Transducer rails
Spring (2pcs)