Foil Test
An aluminum foil test is a method to evaluate the general
cleaning intensity and soundwave coverage throughout the
tank. Perform this test as follows:
Obtain any regular grade (not heavy duty) aluminum foil
sold in a supermarket (less than 0.001 in. thick). Cut a
piece large enough to cover ½ to ¾ of your tank bottom.
Fill the tank as usual with room temperature L&R
General Purpose Cleaner Concentrate (L&R 012) or
L&R Jewelry Cleaner Concentrate, diluted as specified.
Run the machine for 10 minutes without a basket and
with heat off, to de-gas the solution.
Lower the foil into the operating tank vertically and
centered to the bottom.
Continue to hold the foil and let the machine run for one
minute.
Turn the machine off and remove the foil.
Examine your piece of foil. There should be a peening
effect and/or some perforations on the foil. The location
and size of those peening or perforation patterns
indicate the cleaning intensity and uniformity of the
ultrasonic sound waves in the tank.