Temperature
difference
Flow
temperature
Quantity of cold
(only in tariff register 1)
Quantity of heat
(in tariff register 2 and
in current heat quantity)
+ 0,2 K
- 0,2 K
Min. flow temperature for cold quantity sensing
("Cold quantity threshold")
Max. flow temperature for heat quantity sensing
("Heat quantity threshold")
Quantity of heat
(only in current heat quantity)
Display of the
tariff status
:
Temperature difference > 0.2 K and flow
temperature > “heat threshold”
Quantity of heat is accumulated; “bar
below” on LCD
Temperature difference < - 0.2 K and
flow temperature < “cold threshold”
Quantity of cold is accumulated;
“bar above” on LCD
All other situations
“Neither – nor”; all 3 bars
Tariff control by means of tariff timer switch
(tariff “T10”)
Quantity of heat
Volume
TR1
Accumulated measured value:
W or V
W
V
Summation is
performed:
During tariff time
Tariff registers:
Standard registers:
Tariff time
Tariff situation "T10"
Tariff register
Content
TR1
Quantity of heat or volume during tariff time
Tariff time 1
Tariff
OFF
ON
daily
Two settable switching points specify at what time each day the tariff will be activated or
deactivated.