Page 5

User Guide

198.240

2.3

Alignment And Scan Direction

The Land LSP Linescanning Thermometer incorporates an integral laser unit (model dependent) to aid alignment
during installation. Details on the use of this feature are described in section 4.0. Where this laser alignment
feature is not available, precise alignment may be accomplished through the use of a mobile 'hot source', to
identify the plane and extents of the scan line. Alternatively, the intended target product may be used if this is
available during the installation process.

To interpret the scan profile produced by the scanner, it is necessary to determine the scan direction of the unit.
This is indicated by an embossed arrowhead cast into the top of the scanner casing. Once positioned, the design
of the scanner mountings ensure that the unit is replaced in the correct orientation, should the scanner be
removed from the application for servicing etc.

2.4

Installation Precautions

Vibration:

It is highly recommended that the LSP Linescanning Thermometer is positioned in a location where vibration is
kept to a minimum.

Infrared Reflections:

All practical targets have some reflectivity within the infrared waveband. This reflectivity increases as the target
emissivity (see section 5.0) reduces. Therefore the scanner sees some radiant heat energy from reflections of
surrounding objects as well as from the target source. This energy is not usually sufficient to have an effect on
the measurement, but care should be taken to ensure that objects of similar or higher temperature to the target
are not placed where reflected energy may enter the scanner.

Scan

Direction

Arrowhead

Scan

Direction

Product Pass

Direction

Product

LSP

Scanner

Fig. 2 - Scan Direction

240003

Summary of Contents for LSP50

Page 1: ...inst com Internet www landinst com Land Instruments International 2003 Publication N LSPS198 240 Issue A 05 03 Land Instruments International 10 Friends Lane Newtown PA 18940 1804 U S A Telephone 215...

Page 2: ......

Page 3: ...agement System of Land Instruments International Ltd is approved to BS EN ISO9001 2000 for the design and manufacture stockholding in house repair and site servicing of non contact temperature measuri...

Page 4: ...current IEC 617 2 No 02 02 06 Three phase alternating Current IEC 417 No 5017 Earth ground terminal IEC 417 No 5019 Protective conductor terminal IEC 417 No 5020 Frame or chassis terminal IEC 417 No 5...

Page 5: ...SO 3864 No B 3 6 Caution ISO 3864 No B 3 1 WARNING Risk of electric shock BS EN 100015 Observe precautions for handling electrostatic discharge sensitive devices Refer to the operating instructions BS...

Page 6: ...Blank User Guide 198 240...

Page 7: ...allation Precautions 3 0 SERVICES 5 3 1 Electrical Connections 3 2 Water Requirements Water Cooled Units Only 3 3 Air Requirements Air Purged Cooled Units Only 4 0 LASER ALIGNMENT 7 4 1 Using The Lase...

Page 8: ...1 Introduction 8 0 MAINTENANCE 14 8 1 Air Purge Services 8 2 Water And Air Cooling Services 9 0 INSTALLATION ACCESSORIES 15 9 1 Basic Mounting Plate 9 2 Right Angle Mounting Bracket 9 3 Water Cooled...

Page 9: ...n operation and maintenance of the LSP Linescanning Thermometer and should be used in conjunction with the information detailed within the LSP Linescanning Thermometer Installation Guide LSP Processor...

Page 10: ...Blank User Guide 198 240...

Page 11: ...SYSTEM comprises LSP Linescanning Thermometer LSP Processor LSP Interconnection Cable LANDSCAN NT Software Optional LANDSCAN for Windows Configuration Software Standard supplied with LSP Processor or...

Page 12: ...el displays the scanner serial number the model number and the CE conformity mark The model number indicates specific information about the unit including Wavelength and Temperature Range Example Syst...

Page 13: ...100Hz in 10Hz steps TargetWidth Targetdistance 1200mm 47 2in 12mm 0 5in Targetdistance 1200 to 3000mm 47 2 to 118 1in FoV 100 1 Speed Of Response Tau on rising edge linearoutput 7 s AmbientTemperatur...

Page 14: ...capable of carefully following the procedures warnings and guidelines given It is adviseable that all User Guides and Installation Guides associated with LSP System equipment should be read thoroughly...

Page 15: ...e scanner mountings ensure that the unit is replaced in the correct orientation should the scanner be removed from the application for servicing etc 2 4 Installation Precautions Vibration It is highly...

Page 16: ...could result in incorrect operation of the instrument 3 2 Water Requirements Water Cooled Units Only When the LSP Linescanning Thermometer is to be used in an environment where ambient temperatures ar...

Page 17: ...mounting plate If possible ensure that the air purge exit direction is downwards and ensure that this exit is not impeded by mounting structures The air purge flow rate depends upon local conditions...

Page 18: ...arning CLASS 2 Laser Product DO NOT attempt to disassemble the laser unit or any of it s mounting components Embedded laser has a maximum output of 3mW with a beam divergence of 5mrad Caution Caution...

Page 19: ...laser light is projected through the window of the scanner indicating the position and extents of the scan line Adjust the scanner position as required to ensure that the product almost fills the sca...

Page 20: ...Monel oxidised 0 45 Glass 0 90 3mm Nichrome oxidised 0 90 Paper 0 90 Polypropylene METALS Polyethylene 0 90 2mm thickness Aluminium oxidised 0 30 0 96 5mm Copper oxidised 0 85 Iron Steel oxidised 0 85...

Reviews: