Page 7
Risk of damage!
• Make sure that the device is switched off when removing and
installing supplementary modules such as the flow module
(this means that the 10 V (21) and 5 V (20) LEDs must be off
when a module is removed or installed)!
• Make sure that module polarity is correct when the module is
inserted!
• Make sure that the spacers snap into place.
• Make sure that all wire inlets are sealed with the supplied screw connections
(
24
).
• Only use wire that is 3 to 6.5 mm in diameter.
• To ensure a leakproof seal, avoid using twisted wires.
• Seal all unused wire inlets with the supplied dummy plugs (
25
).
Risk of damage!
Only the supplied hose connectors are to be used.
The use of angled or narrow connections could result in false flow
readings (owing to a nozzle effect).
• Make sure that the flow direction is adhered to!
• Only dry, oil-free, solvent-free and filtered (5 µm) air should be conducted into
the device!
• Do not make any modifications to the device!!
• Mounting, maintenance and repair work should only be performed by trained
personnel!
• Make sure the device is attached to the wall with the wire connectors facing
upwards.
• Only use original LANCIER replacement parts!
Important!
Obey handling instructions. Electrostatic discharge (ESD) damage.
IMPORTANT!
Observe the EMC directives!
The screen of screened cables must be connected to the marked
ground clamps.
Summary of Contents for ACS 10000