20
Biguanide Plus Borate Disk
(Code 4331)
Biguanide
Biguanide Shock
Total Alkalinity
pH
Calcium Hardness
Copper
Iron
Borate
pH
Biguanide
Shock
Biguanide
Blank
Iron
Copper
Calcium
Hardness
Total
Alkalinity
Borate
MULTIPLE-USE DISKS
Chlorine/Bromine/pH
(Code 4334)
pH
Free Chlorine
Bromine
Total Chlorine
pH
Free
Chlorine/
Bromine
Total
Chlorine
Blank
pH
pH
Blank
Blank
Free
Chlorine/
Bromine
Free
Chlorine/
Bromine
Total
Chlorine
Total
Chlorine
Chlorine/Bromine/pH/Alkalinity
(Code 4335)
pH
Free Chlorine
Bromine
Alkalinity
pH
Free
Chlorine/
Bromine
Alkalinity
Blank
pH
pH
Blank
Blank
Free
Chlorine/
Bromine
Free
Chlorine/
Bromine
Alkalinity
Alkalinity
MAINTENANCE
CLEANING
The optical system of the WaterLink
®
Spin Touch
®
must be kept clean and dry for optimal
performance. Dry the disk with a lint-free wipe before placing it into the chamber to avoid
introducing moisture. For best results, store the instrument in an area that is dry and free
from aggressive chemical vapors. Clean the exterior housing with a damp, lint-free cloth.
Do not allow water to enter the light chamber or any other parts of the meter. To clean the
light chamber and optic lenses, point a can of compressed air into the light chamber and
the lid and blow the pressurized air into the light chamber and lid. Focus the pressurized
air around the LEDs which are the small round lenses positioned at 2:00, 4:00, 6:00, 8:00,
10:00 and 12:00 in the lid. The photodiodes are located on the bottom of the chamber
around the hub. This area must be kept clean and dry. Use a cotton swab dampened with
Windex
®
window cleaner to gently swab the LED and photodiode lenses. Do not use alcohol;