12
LaMotte TRACER pH PockeTester
TM
1741 Instruction
4. Replace the battery compartment cap.
NOTE: If the batteries are removed, stored data and user calibrations will
be cleared. A new user calibration will be required. Factory calibrations will
be retained.
All EU users are legally bound by the Battery Ordinance to return all
used batteries to community collection points or wherever batteries/
accumulators are sold.
Disposal in household trash or refuse is prohibited.
DISPOSAL: Follow the valid legal stipulations in respect of the disposal of the
device at the end of its life cycle
Other Battery Safety Reminders
•
Never dispose of batteries in a fi re. Batteries may explode or leak.
•
Never mix battery types. Always install new batteries of the same type.
Cleaning and Conditioning the Electrode
Surface Cleaning – Only in case of visible surface contamination or if readings
should become erratic, use a disposable laboratory towel wetted with ethanol or
isopropyl alcohol and gently rub the surface, until no more residue is visible.
Replacing the Electrode
The TRACER is shipped with an electrode attached. Electrode life is limited
and is dependant on the frequency of use and care. If the electrode needs to be
replaced, follow these steps for removing and connecting electrodes.
1. Turn meter off.
2. To remove the electrode, turn the collar counter-clockwise and remove it.
3. Gently rock the electrode from side to side, pulling it downwards, until it
disconnects from the meter.
4. To attach an electrode, carefully plug the electrode into the meter socket.
Note that the electrode connector is keyed to ensure a proper connection.
CAUTION: Take case to align pins carefully. Bent or broken ins will cause
the meter to malfunction.
5. Tighten the electrode collar fi rmly enough to make a good seal. A rubber
gasket will seal the electrode with the meter.