![LaMotte TRACER pH PockeTester Instruction Download Page 11](http://html1.mh-extra.com/html/lamotte/tracer-ph-pocketester/tracer-ph-pocketester_instruction_3427743011.webp)
LaMotte TRACER pH PockeTester
TM
1741 Instruction
11
Recalling Stored Readings
NOTE: First ensure that the
HOLD
symbol is not displayed. If it is, exit the
MODE function by pressing the MODE button.
1. Press the CAL button once and then press the MODE button
immediately
after
CAL
is displayed; the location number (1 through 15) will fl ash. If the
CAL mode is accidentally accessed (display fl ashing), press the CAL button
again to exit.
2. The last stored reading taken will be displayed fi rst. To advance through the
stored readings, press the MODE button. The location number is displayed
fi rst, followed by the reading stored in that location.
3. To exit the recall mode, press the CAL button and the TRACER will return to
normal operation. If the batteries are removed, any stored readings and user
calibrations will be discarded.
MAINTENANCE
Care/Storage of Electrode
1. Always rinse the electrode in tap, distilled or deionized water between
measurements to avoid cross-contamination of the samples. Double rinsing is
recommended when high accuracy is required.
2. To preserve electrode life
keep the sponge in the protective cap soaked
with a pH 4 buffer.
Cap TRACER when not in use.
Store vertically
.
Do Not allow sensor to dry out.
3. Do not touch the electrodes. Touching the surface of the electrodes may damage
and reduce the life of the electrodes.
4. Salt deposits may build up in the storage cap and should periodically be
rinsed away. These deposits could affect measurements at low conductivity.
Battery Replacement
1. Twist off the battery compartment cap.
2. Hold the battery housing in place with one fi nger. Remove the battery
carrier by pulling on the small tabs.
3. Replace the four CR2032 batteries. Observe polarity.