8
6. Insert tube into chamber, close
lid and select
Scan Blank
.
7. Remove tube from colorimeter.
Add 5 drops of *Copper 1
(6446).
8. Cap and invert to mix. Wipe tube
dry.
9. Insert tube into chamber. Close
lid. Select
Scan Sample
. Record
result as Copper.
036 Cu Thiocarbamate
0.99
ppm
Scan BLank
Scan Sample
12:00:00 001/500
NOTE: The meter will remember the last scanned blank reading. It is not
necessary to scan a blank each time the test is performed. To use the
previous blank reading, instead of scanning a new one, scroll to Scan Sample
and proceed. For the most accurate results, the meter should be blanked
before each test and the same tube should be used for the blank and the
reacted sample.
NOTE: For the best possible results, carry a reagent blank through the
procedure. After scanning the blank, perform the test procedure on clear,
colorless, distilled or deioized water. Subtract results of regent blank from all
subsequent test results.
Analysis