11
10. Press
ENTER
to select
Calibrate
.
A reverse font (light background
with dark characters) will appear
to indicate that the reading can
be adjusted.
036 Cu Thiocarbamate
0.99
ppm
,
=Edit, ENTER=Save
+ENTER=Default
12:00:00 001/500
11. Press
or
to
adjust the value shown to the
concentration of the prepared
standard, 1.00 in this example.
NOTE: A maximum adjustment
of 25% is possible.
036 Cu Thiocarbamate
1.00
ppm
,
=Edit, ENTER=Save
+ENTER=Default
12:00:00 001/500
12. Press
ENTER
to save the value.
To leave the Calibration
procedure without saving the
adjustment, press
EXIT
.
Press
and
ENTER
at any
time to return to the default
value.
The calibration has now been
standardized and can be used
for testing. Scroll to
Scan Blank
and begin testing.
036 Cu Thiocarbamate
1.00
ppm
%T/Abs
Calibrate
12:00:00 001/500