10
17. At the end of the 12 minute
waiting period, invert once to
mix. Immediately insert tube into
chamber. Close lid. Select
Scan
Sample
. Record result as Urea.
096 Urea
3.3
mg/L
Scan Blank
Scan Sample
12:00:00 001/500
NOTE: The meter will remember the last scanned blank reading. It is not
necessary to scan a blank each time the test is performed. To use the previous
blank reading, instead of scanning a new one, scroll to Scan Sample and
proceed. For the most accurate results, the meter should be blanked before
each test and the same tube should be used for the blank and the reacted
sample.
NOTE: For the best possible results, carry a reagent blank through the
procedure. After scanning the blank, perform the test procedure on clear,
colorless, distilled or deioized water. Subtract results of regent blank from all
subsequent test results.
Analy
sis