background image

3

DEVICE CONNECTION

The WaterLink

®

 Spin Touch

®

 supports connections to a Windows

®

 based PC (over USB) and 

to Android and iOS mobile devices (via Bluetooth).

CONNECTING VIA USB 

Using the provided USB cable, the WaterLink

®

 Spin Touch

® 

can be connected to a Windows

® 

based PC. Before connecting a meter via USB, download and install the free WaterLink

®

 

Connect Windows

®

 application from 

http://softwarecenter.lamotte.com

. When the 

WaterLink

®

 Spin Touch

®

 is connected to the PC via USB, the onboard touch screen controls 

will be disabled and operation of the meter is performed using the WaterLink

®

 Connect 

application. LaMotte off ers robust water analysis programs as well, such as WaterLink

®

 

Solutions

, to collect test results and off er detailed treatment recommendations. Learn 

more about LaMotte software products at 

http://softwarecenter.lamotte.com

.

CONNECTING VIA BLUETOOTH

The WaterLink

®

 Spin Touch

®

 is capable of connecting to a Bluetooth enabled device, such 

as a phone or tablet. The Spin Touch is also compatible with a LaMotte Bluetooth Printer 
[Code 5-0066]. Other Bluetooth printers are not supported.

Transferring results over Bluetooth to a mobile device requires that a mobile app is 
installed on the device and an active account for the associated software program. For 
example, with an active WaterLink

®

 Solutions

 account and the WaterLink

®

 Solutions

 

mobile app on a phone or tablet, results can be transferred from the meter into the 
application. Mobile apps for LaMotte software products are available from iTunes (for 
iOS devices) and Google Play (for Android devices). Learn more about LaMotte software 
products at 

http://softwarecenter.lamotte.com

.

To transfer results from the WaterLink

®

 Spin Touch

®

 to a LaMotte software product mobile 

app:

1.  Log into the LaMotte software product mobile app.

2.  Search for a customer or Site record. You may need to create one before you can 

begin testing.

3.  Start a Water Test in the mobile app.

4.  Perform a water test from the Spin Touch onboard touchscreen normally. The Spin 

Touch and the Bluetooth enabled mobile device will connect automatically. 

5.  When the connection is available the 

 will light on the touch screen. When the 

 

is dim, the meter and device are not connected. Tap the 

 to transfer results to the 

mobile app.

Whenever the 

  appears on the touch screen the Spin Touch is capable of connecting 

to the Mobile Bluetooth Printer (Code 5-0066). The 

 button will be highlighted when 

connected and dim when not connected. 

The Spin Touch cannot connect to both mobile device and the printer at one time. (See the 
Save, Print and Send, page 7)

USING BLUETOOTH LOW ENERGY

The WaterLink

®

 Spin Touch

®

 will automatically connect to the nearest Bluetooth enabled 

phone or tablet or printer. It is not necessary to pair the Spin Touch to the Bluetooth 
enabled device. Barriers to wireless signals can reduce the range of wireless devices. 
The WaterLink Spin Touch will work best if there are no walls between it and the receiving 
devices.

DEVICE CONNECTION   

Summary of Contents for 3587

Page 1: ...WaterLink n F Touch F Code 3587 Instruction Manual ...

Page 2: ... will require updates To do so 1 Visit http softwarecenter lamotte com and download the WaterLink Connect application for Windows 2 Plug the Spin Touch into the computer using the included USB cable 3 Open the WaterLink Connect desktop application and wait for the update to complete Once the update is complete it is safe to close WaterLink Connect and unplug the lab After updating firmware it is re...

Page 3: ...n be transferred from the meter into the application Mobile apps for LaMotte software products are available from iTunes for iOS devices and Google Play for Android devices Learn more about LaMotte software products at http softwarecenter lamotte com To transfer results from the WaterLink Spin Touch to a LaMotte software product mobile app 1 Log into the LaMotte software product mobile app 2 Searc...

Page 4: ...e 2 Do not overfill the disk If the disk is overfilled sample water will flow out of the overflow hole in the center of the disk The disk is not leaking Dry the disk and run the test Overfilled Dry disk before placing in meter 3 Do not under fill the disk If the disk is under filled the reagent chambers will not fill entirely and results will be inaccurate Underfilled 4 Do not introduce air bubbles into ...

Page 5: ... in progress and the disk is spinning Do not open the lid when the disk is spinning Care should be taken when closing the lid Do not slam the lid Wiring between the lid and the body of the photometer passes through the hinge The meter will not run with the lid open TOUCHSCREEN DISPLAY When the button is pressed to turn the meter on the Test Screen will be displayed 12 14 PM 30 Jan 19 Water type Di...

Page 6: ... test results to the test log if Auto Save is not enabled Tap the hightlighted to send the results to the enabled Mobile Bluetooth Printer Tap the highlighted to send the results to a Bluetooth enabled device Tap to return to the Test Screen 10 Press and hold for 2 seconds to turn the meter off For the most accurate results samples should be at room temperature The blank well allows the readings to...

Page 7: ...ple at a time Auto Save Tests must be disabled If Auto Save Tests is disabled will be highlighted on the Test Results screen After the test has been run tap to save the results for that water sample to the Test History A user defined identification tag can be assigned to an individual sample result The tag can be up to 4 hexadecimal characters It cannot begin with a zero To tag a test result tap on ...

Page 8: ...ge fits into the fill hole on the disk The syringe tip should not be removed from the syringe Syringes should be cleaned between water samples Pump air in and out of the syringe a few times to clear the previous sample or rinse the syringe with a small amount of the next water sample before filling it with the next sample REAGENT DISK The WaterLink Spin Touch uses a disk reagent system The dried reag...

Page 9: ...ch to a Windows based PC When used with the AC Power Adapter it connects the meter to an AC outlet WARNING only use the USB cable and wall adapter that are supplied with the kit Make no substitutions BATTERY A fully charged battery will last for approximately 150 tests under average conditions The battery life will vary based on usage patterns The meter should be turned off after testing to prolong...

Page 10: ...ment of the hub and disk Tap to exit to the Settings menu Power Options There are three power options Auto Dim Time Auto Off Time and Power Tap the options then tap a selection Tap selection Tap to exit to the Settings menu Bluetooth There are two Bluetooth options Bluetooth Enabled and Fast Printer Connect that are used to print test results Tap to exit to the Settings menu Other Settings About li...

Page 11: ...of range of the reagent system wil be RED RED test results may not be accurate Go SETTINGS Other Settings to turn the Ranges Enabled feature OFF Testing samples for Nitrate immediately after treatment with a chlorine neutralizer containing sodium thiosulfate will give low results Retest in 2 3 days If the concentration for one test factor is significantly out of the range for the reagent system the...

Page 12: ...Range pH Blank Low Range Hardness High Range pH Empty Nitrate Phosphate Ammonia Fish Farming Salt Water Disk FF203 Code 4352 pH Phosphate Calcium Magnesium Alkalinity Nitrite Nitrate Ammonia pH Nitrite Alkalinity High Range pH Blank Magnesium Phosphate Empty Nitrate Calcium Ammonia NOTE Colored reagents may be visible in the disk before adding sample water ...

Page 13: ...terLink Spin Touch FF reagent disks see page 12 SPECIFICATIONS Instrument Type Centrifugal Fluidics Photometer Wavelengths interference filters 390 nm 428 nm 470 nm 525 nm 568 nm 635 nm Display Color Capacitive Touchscreen 3 5 in 320 x 240 pixel resolution Wavelength Accuracy 2 nm Wavelength Bandwidth 10 typical Photometric Range 2 to 2 AU Photometric Precision 0 01 AU at 1 0 AU Photometric Accurac...

Page 14: ...ater Resistance Rubber over molded base rubber USB Port Plug gasketed display and hinge Electrical Rating Rated voltage 5V Rated power of input current 1 6 A at USB C Auto Off Yes default 15 only with battery power Power Save Yes default OFF Data Logger 250 test results stored for download to PC or transfer by Bluetooth Certifications EZ BLE PRoC Module CYBLE 022001 00 RF Radio FCC USA Industry Cana...

Page 15: ...ill disk correctly see FILLING Range Error Raw data out of range Contact Support Output Error Decreased light intensity Possibly dirty lens Clean lens see CLEANING Follow Range Check Procedure If error message persists contact Support Consistently unexpected high results for metals Metals may actually be present Repeat test with distilled water If the results still show that metals are present con...

Page 16: ...er not on The Spin Touch will only print to the Mobile Bluetooth Printer Code 5 0066 Turn printer on Connect to Mobile Bluetooth Printer Code 5 0066 TROUBLESHOOTING WITH THE METER CHECK DISK Do not attempt to separate the components of the Meter Check Disk Code 1705 The Meter Check Disk consists of a disk with a permanently attached cover Do not fill the Meter Check Disk with water Water is not us...

Page 17: ...erforming satisfactorily Angle Calibration Successful will be displayed and the settings will be saved If the meter is not performing satisfactorily Angle Calibration Unsuccessful Contact Tech Support will be displayed 8 Tap g to return to the Testing Menu and resume testing email softwaresupport lamotte com phone 800 344 3100 option 3 Mon Fri 9am 5pm EST HELPFUL HINTS Do not touch top or bottom o...

Page 18: ...mber must be obtained from LaMotte Company by calling 800 344 3100 ext 3 US only or 410 778 3100 ext 3 faxing 410 778 6394 or emailing softwaresupport lamotte com Often a problem can be resolved over the phone or by email If a return of the meter is necessary attach a letter with the return authorization number meter serial number a brief description of problem and contact information including ph...

Page 19: ...ents is available 24 hours a day from the National Poison Control Center at 1 800 222 1222 or by contacting the 24 hour emergency line for ChemTel at 1 800 255 3924 USA Canada Puerto Rico For locations outside of the North American continent call 813 248 0585 collect Ensure that the protection provided by this equipment is not impaired Do not install or use this equipment in a manner that is not i...

Page 20: ... defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification LaMotte Company specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct indirect incidental or consequential damages LaMotte Company s total liability is l...

Reviews: