TROUBLESHOOTING GUIDE
Error Messages
Err1
Very Low battery. Replace battery or switch to AC power. Press back
arrow (
) to back out. Scan sample again. Replace battery as soon
as possible.
Err2
The meter can not be calibrated outside of the allowable range of
the displayed reading. Confirm that standard was made correctly.
The displayed reading can only be adjusted to ±50% of the factory
calibration.
Err3
Meter can not be calibrated with a zero sample. Calibrate the meter
with a sample other than zero.
Err4
Processing error due to motion of suspended particles or submicron
air bubbles or opening/closing lid during readings. Scan sample
again until reading is obtained.
Err5
No blank reading. The meter has never been blanked for this test
factor. Blank meter.
Err6
Internal mathematical error. Re-blank the meter and rescan the
sample.
Err7
Configuration error. Call LaMotte Tech Service. Meter may have to
be returned for repairs.
low battery
Low battery. Change battery.
Troubleshooting
PROBLEM
REASON
SOLUTION
Lost in meter menus
Reset to factory default
settings. (see page 8)
Turn meter off. Hold down
*
I
OK
and press
ON
.
Release both buttons. Press
*
I
OK
to select the default
settings. Meter will turn off
and the factory settings will
be restored.
Usually large negative
or positive readings
when performing
calibration.
Incorrect standards used to
calibrate meter.
Use fresh 0.0 standard in
clean, labeled, vial. Reset
meter to factory default
settings. See procedure
above. Recalibrate meter.
Erroneous readings
Measurement was taken
with lid open.
Close lid. Read again.
>
on display
Over range.
The sample is outside of the
acceptable range. Dilute
sample and test again.
30
TROUBLESHOOTING