The Japan Standard calibration mode should be used only with Japanese
Polystyrene Standards (0–100 NTU).
Note: StablCal
®
standards below 50 NTU should not be used to calibrate the
2020. The diluent has a different refractive index than traditional formazin
standards and will affect the results.
6. Press
to exit to a previous menu or
make another menu selection or
press
OFF
to turn the meter off.
AVERAGING
The averaging option allows the user to average multiple readings. This option will
improve the accuracy of samples with readings that may tend to drift with time. When
the two, five or ten reading options have been selected, the meter will show a running
average of the readings that have been taken until the final average is displayed. The
default setting is disabled. To change the setting:
1. Press
ON
to turn the meter on.
1 . 3
2. Scroll down and then press
*
I
OK
to select
Options
.
M a i n M e n u
M e a s u r e
D a t a L o g g i n g
* O p t i o n s
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
3. Press
*
I
OK
to select
Averaging
.
O p t i o n s
* A v e r a g i n g
T u r b i d i t y
D a t e / T i m e
L a n g u a g e
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
OPTIONS & SETUP • AVERAGING