Troubleshooting

Problem

Cause

Action

Power on but no
display

Batteries

Insert batteries

Batteries

Verify correct polarity

Batteries

Replace batteries

Unstable readings

Electrode

Immerse electrode more deeply in
sample

Electrode

Condition electrode before first use.
Soak in tap water for 10 minutes.

Electrode

Remove air bubbles caught under
electrode

Electrode

Clean electrode

Electrode

Replace electrode

Static charge

Wiping electrode Rinse electrode and blot. Do not wipe

electrode

Same readings in
different samples

Cracked or broken
electrode

Replace electrode. Avoid plunging
electrode into bottom of container and
stir bars.

Display frozen

HOLD function

Press MODE or turn meter off

Button response

Remove batteries (stored data will be
lost)

Steady “

-1

” display Wait

Reading not stable yet

Replacing the Electrode

The TRACER is shipped with an electrode attached. If the electrode
needs to be replaced, follow these steps for removing and connecting
electrodes.

1.

Turn meter off.

2.

To remove the electrode, turn the collar counter-clockwise and
remove it.

3.

Gently rock the electrode from side to side, pulling it downwards,
until it disconnects from the meter.

4.

To attach an electrode, carefully plug the electrode into the meter
socket. Note that the electrode connector is keyed to ensure a
proper connection.

5.

Tighten the electrode collar firmly enough to make a good seal. A
rubber gasket will seal the electrode with the meter.

9

Summary of Contents for 1742 PockeTester

Page 1: ...TRACER POCKETESTER ORP CODE 1742...

Page 2: ......

Page 3: ...asic Operation Powering the TRACER 6 Electrode Recognition 6 Automatic Calibration 6 Data Hold 6 ORP Testing Overview 6 ORP Display 6 Getting Started 6 ORP Measurement 6 Storing Readings 7 Recalling S...

Page 4: ...unction LCD with Bar graph Operating Conditions 32 to 122 F 0 to 50 C and 80 RH ORP Range Accuracy 999mV 4mV Measurement storage 15 tagged numbered readings Power Four CR2032 button batteries see Page...

Page 5: ...Battery compartment cap 2 LCD Display 3 MODE HOLD button 4 CAL RECALL button 5 ON OFF button 6 Electrode Collar 7 Electrode Note The Electrode cap is not shown Display Description 1 Reading 2 Units 3...

Page 6: ...Hold Press the MODE HOLD button to freeze the current reading The HOLD display icon will appear along with the held reading The held reading will be stored in the next 15 reading storage location Pre...

Page 7: ...epeat step 1 to store the next reading 4 If an attempt is made to store more than 15 readings the stored readings starting with the first reading will be overwritten Recalling Stored Readings NOTE Fir...

Page 8: ...tore vertically Battery Replacement 1 Twist off the battery compartment cap 2 Hold the battery housing in place with one finger Remove the battery carrier by pulling on the small tabs 3 Replace the fo...

Page 9: ...r and stir bars Display frozen HOLD function Press MODE or turn meter off Button response Remove batteries stored data will be lost Steady 1 display Wait Reading not stable yet Replacing the Electrode...

Page 10: ...of one 1 year from the original purchase date The probe is guaranteed to be free from defects in material and workmanship for a period of six 6 months from the original purchase date In the event that...

Page 11: ...LaMotte Company will service and repair out of warranty products at a nominal charge...

Page 12: ...aMOTTE COMPANY Helping People Solve Analytical Challenges PO Box 329 Chestertown Maryland 21620 USA 800 344 3100 410 778 3100 Outside U S A Fax 410 778 6394 Visit us on the web at www lamotte com 5 08...

Reviews: