When the unit is connected to a DMM which accuracy is
supposed to be 0.5% in 200mV range,the reading displayed
on the DMM will be 10.25mV max. (10.20mV x (1+0.5%)=10.
25mV) and 9.75mV min. (9.80mV x (1-0.5%)=9.75mV)
Current Clamp Accuracy
DCA range: 1mV/10mA
± (2.0%±5mA) 10mA~20A
DCA range: 1mV/100mA
± (2.0%±20mA) 100mA~40A
± (4.0%±0.3A) 40A~60A
ACA range: 1mV/10mA
±(2.0%±5mA) 10mA~10A (40Hz~2kHz)
±(4.0%±30mA) 10mA~10A (2kHz~10kHz)
±(6.0%±30mA) 10mA~10A (10kHz~20kHz)
±(8.0%±30mA) 10A~15A (40Hz~20kHz)
ACA range: 1mV/100mA
±(2.0%±30mA) 100mA~40A (40Hz~1kHz)
±(4.0%±30mA) 100mA~40A (1kHz~2kHz)
±(6.0%±30mA) 100mA~40A (3kHz~5kHz)
±(8.0%±0.3A) 40A~60A (40Hz~5kHz)
Load Resistance: 10k
Ω
typical
Temperature Coefficient: 0.1x(specified accuracy) per
degree C. (0°C
to 18°C, 28°C to 40°C)
SAFETY INFORMATION
Cleaning
Wipe the case with a damp cloth and mild detergent. Do not
use abrosives or solvents. Dirt or moisture in the terminals
can affect readings.
S a f e t y:
Conforms to EN 61010-1, EN 61010-2-032,
CAT II 300V, Class II, Pollution degree 2 Indoor use.
CAT II:
Is for measurements performed to the low-
voltage on circuits directly connected.
EMC:
Conforms to EN 61326-1.
The symbols used on this instrument are:
Dangerous voltage.
Caution, refer to accompanying documents
Equipment protected throughout by Double
insulation (Class II)
Alternating current
Direct current
AC/DC
mA CURRENT CLAMP
OPERATOR’S MANUAL
CLAMP
300V
60A
CAT.II
7000-1698A