![Lake Shore 370 User Manual Download Page 168](http://html.mh-extra.com/html/lake-shore/370/370_user-manual_670632168.webp)
Lake Shore Model 370 AC Resistance Bridge User’s Manual
Model 370 Resistance Input Calibration Work Sheet for Paragraph 8.12.2.6 (Continued)
Step Range/Setting Coefficient
# Resistor Voltage Filter
Actual
RDGR?
Coefficient
6 a
316 nA Gain
CALG 1,12
200 k
Ω
63.2
mV 10s
6 b 100 nA Gain
CALG 1,11
200 k
Ω
20
mV 10s
6 c
31.6 nA Gain
CALG 1,10
200 k
Ω
6.32
mV 10s
6 d 10 nA Gain
CALG 1,9
200 k
Ω
2
mV 10s
6 e
3.16 nA Gain
CALG 1,8
200 k
Ω
632
μ
V
10s
6 f
1 nA Gain
CALG 1,7
200 k
Ω
200
μ
V
10s
6 g 316 pA Gain
CALG 1,6
200 k
Ω
63.2
μ
V
50s
6 h 100 pA Gain
CALG 1,5
200 k
Ω
20
μ
V
50s
6 i
31.6 pA Gain
CALG 1,4
200 k
Ω
6.32
μ
V 100s
6 j
10 pA Gain
CALG 1,3
200 k
Ω
2
μ
V 200s
Current Gain Coefficient = Actual Resistance /RDGR
Step Range/Setting Coefficient
# Resistor Voltage Filter
Actual
RDGR?
Coefficient
7a
3.16 pA Gain
CALG 1,2
600 k
Ω
2
μ
V 200s
7b
1 pA Gain
CALG 1,1
2 M
Ω
2
μ
V 200s
Voltage Gain Coefficient = Actual Resistance/RDGR
8.12.3
Analog Voltage Output Calibration
The Model 370 has 2 analog voltage outputs with a range of
±
10V. These outputs can be calibrated independently of
each other and their calibration is independent of the measurement input. A DC voltmeter with accuracy of 1 mV on a
10 V range and cable to connect to the analog output are required. A zero offset and gain coefficient must be calculated
for each output. The method used for calculating these coefficients provides the most accurate calibration of the DAC
used in the instrument.
To calibrate the analog voltage outputs perform the following sequence of operations.
1. Configure the analog voltage output for manual operation (Paragraph 4.16). Set the manual output value to +100%.
Attach the voltmeter and allow the instrument to warm up for 1 hour.
2. Set the zero offset and gain calibration coefficients to their defaults:
Analog Output 1:
CALZ 4,1,0
CALG 4,1,1
Analog Output 2:
CALZ 4,2,0
CALG 4,2,1
3. Measure and record the +100% output voltage.
4. Set the manual output value to –100%.
5. Measure and record the –100% voltage.
6. Calculate the gain coefficient as: Coefficient = 20 V/(Voltage (+100%) – Voltage (–100%)).
The gain should be 0.995 ±1%.
7. Send the gain coefficient for the appropriate output to the instrument:
Analog Output 1:
CALG 4,1,Coefficient
Analog Output 2:
CALG 4,2,Coefficient
8. Set the manual output value to 0%.
9. Measure and record the output voltage.
8-22
Service