background image

Instructions,  
Platinum Resistance Thermometer Installation 

 

 

Lake Shore Cryotronics, 575 McCorkle Blvd. Westerville, OH 43082 

www.lakeshore.com

  

Phone (614) 891-2243 – Fax (614) 891-1362 – 

[email protected]

 – 

[email protected]

 

Model PT-102, PT-103, and PT-111 

The PT-102, PT-103, and PT-111 are wire-wound thermometers and normally used as 
secondary standards. However, these thermometers should be treated in the same manner 
as any precision instrument. It is recommended that they not be subjected to any 
unnecessary shock or rough mechanical treatment. 

 

In reading the thermometer, it is important to use the proper measuring current. An excessively 
high current will cause joule heating (I

2

R), thus giving erroneous reading. Lake Shore 

recommends a maximum power dissipation (joule heating) of 10 µW: the same as used in  
calibration. The recommended measuring currents to avoid joule heating are as follows: 

 

Ω

 to 200 

Ω

 

0.5 mA 

200 

Ω

 to 1200 

Ω

 

0.1 mA 

 

There are three aspects of using a cryogenic temperature sensor which are critical to its optimum performance: 1) proper 
mounting of the sensor package, 2) proper joining of sensor lead wires and connecting wires, and 3) proper thermal anchoring 
of the connecting wires. Although the sequence in which these concerns should be addressed is not fixed, all elements 
covered under each aspect should be adhered to for maximum operating capabilities of the sensor.
 

MOUNTING ADAPTERS 

PT-102-AL and PT-103-AM each include a platinum RTD sensor mounted into a flat aluminum block. They can be mounted to 
any flat surface with a 6-32 or M3 screw (not included) and Inconel® Belleville washer (included). 

 
Adapter material:

 6061 Al block (PT mounted to adapter using Cotronics Durabond® 950 Al-based adhesive)  

AL leads:

 Two 0.010-inch diameter: 10.160 ±1.270 mm (0.400 ±0.050 in) long  

AM leads:

 Two 0.010-inch diameter: 15.240 ±1.270 mm (0.600 ±0.050 in) long  

Lead material:

 Platinum  

Mass: 

PT-102-AL: 3.8 g, PT-103-AM: 2.1 g  

Limitation:

 The aluminum alloy limits the upper useful temperature of these configurations to 800 K 

 

PT-102-AL 

PT-103-AM 

 

Summary of Contents for PT-102

Page 1: ... 1200 Ω 0 1 mA There are three aspects of using a cryogenic temperature sensor which are critical to its optimum performance 1 proper mounting of the sensor package 2 proper joining of sensor lead wires and connecting wires and 3 proper thermal anchoring of the connecting wires Although the sequence in which these concerns should be addressed is not fixed all elements covered under each aspect sho...

Page 2: ...sh and apply a thin layer on the mounting surface Press the sensor firmly against the varnish during curing to ensure a thin bond layer and good thermal contact Varnish will air dry in 5 to 10 min Sufficient time must be allowed for the solvents in the varnish to evaporate There is a small probability of ionic shunting across the sensor during the cure period of the varnish typically 12 to 24 h St...

Page 3: ...sistance to chemical solvents and burnout Formvar has a temperature rating of 373 K while polyimide has a temperature rating of 505 K 4 For use above 500 K use high temperature connecting wire e g nickel with high temperature insulation e g wrap around ceramic tube or ceramic beaded Connection should be made by spot welding or brazing 5 Platinum sensor leads can easily be soldered if a flux is use...

Page 4: ...ionally the lead material is not pure platinum but rather an alloy or coated wire used to increase solderability and provide a more robust lead The lead material is Pt Rh 5 alloy wire for the PT 103 platinum coated palladium wire for the PT 102 and platinum coated nickel wire for the PT 111 During calibration at Lake Shore contact to each lead is made within 1 mm of the lead end An estimate of the...

Reviews: