background image

enough slack to allow for the thermal contractions that occur during cooling which could fracture a solder joint or lead. 
Insulating the soldering joint is recommended. This can be achieved with heat shrink tubing. 

 

NOTE: 

Be sure when soldering, or any time during use, the package does not exceed 420 K (146 °C). Exceeding this 

temperature may cause a shift in the sensor values. 
 

 

HEAT SINKING/THERMAL ANCHORING 

Since the area being measured is read through the base of the sensor, heat flow through 
the connecting leads can create an offset between the sensor chip and the true sample temperature. Thermal anchoring of the 
connecting wires is necessary to assure that the sensor and the leads are at the same temperature as the sample. 

 

1.  Connecting wires should be thermally anchored at several temperatures between room temperature and cryogenic 

temperatures to guarantee that heat is not being conducted through the leads to the sensing element. The leads of the 
BO package are thermally anchored to the mounting block with a beryllium oxide heat sink chip. 

 

2.  If the sensing leads or connecting wires have a thin insulation such as Formvar™ or Polyimide, a simple thermal anchor 

can be made by winding the wires around a copper post, bobbin, or other thermal mass. A minimum of five wraps around 
the thermal mass should provide sufficient thermal anchoring. However, if space permits, additional wraps are 
recommended for good measure. To maintain good electrical isolation over many thermal cycles, it is good practice to first 
varnish a single layer of cigarette paper to the anchored area then wrap the wire around the paper and bond in place with 
a thin layer of IMI 7031 varnish. Formvar™ wiring insulation has a tendency to craze with the application of IMI varnish. 
Once IMI varnish is applied, the wires cannot be disturbed until all solvents have evaporated and the varnish has fully 
cured (typically 12 to 24 h). 

 

3.  As a minimum, one of the thermal anchors should be as close as possible to the sample itself to ensure thermal 

equilibrium between the sample and temperature sensor. 

 

CRYOGENIC ACCESSORIES

 – Recommended for proper installation and use of CX-10XX-BO sensor: 

Stycast Epoxy

®

 2850FT

 (P/N 9003-020, 9003-021): Permanent attachment, excellent low temperature properties, poor 

electrical conductor, and low cure shrinkage. 

Apiezon

®

 N Grease

 (P/N 9004-020): Low viscosity, easy to use, solidifies at cryogenic temperatures, excellent lubricant. 

IMI 7031 Varnish

 (P/N 9009-002): Nonpermanent attachment, excellent thermal conductor, easy to apply and remove. 

Indium Solder 

(P/N 9007-002-05): 99.99% pure, excellent electroplating material, foil form. 

90% Pb 10% Sn Solder

 (P/N 9008-001): Greater lead content, for higher temperature applications no greater than 200 °C. 

RMA Soldering Flux

 (P/N 9008): Variety of types, refer to Lake Shore Product Catalog for details. 

Phosphor-bronze Wire

 (P/N 9001-00X): Available in single, duo, and quad strands, non-magnetic, low thermal conduction. 

Manganin Wire

 (P/N 9001-00X): Low thermal conductivity, high resistivity, non-magnetic. 

Heat Sink Bobbin

 (P/N 9007-900 Large, 9007-901 Small): Gold plated oxygen-free high-conductivity (OFHC) copper bobbins. 

Instruments: 

Lake Shore sells a complete line of instrumentation used with the CX-10XX-BO sensors, such as current 

sources, cryopump monitors, temperature controllers, monitors and thermometers, temperature scanners and transmitters. 

 
For complete product description and detailed specifications on the above accessories and instruments, consult the 
Lake Shore Temperature Measurement and Control Catalog, call at (614) 891-2243, E-mail at 

[email protected]

,  

or visit our website at 

www.lakeshore.com

 
 
 
 
 
 
 
 
 
 
 
 
 