LAFOMED LFSS12AA Instruction Manual Download Page 19

                                                                           

                                                       

 

  16 

working. It will absolutely keep the user safe and display the error code( See the below page 14). 

If alarm appears, please write down the error code No. and cut off the power. D

on’t open the door and then turn on the 

power again to wait the pressure turn back to “

-0.5~

0.5”.

 

 

 

 

 

Below  is  a  list  of  alarms.  If  the  user  cannot  solve  the  problem  according  to  below  chart,  please  contact  your  local 
distributor or us, we will help the user solve problems. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8. Maintenance 

 

We suggest running program once more to see if the error happens again.

No.

Code

Sound

Conditions

Reason

Resolution

1

E31

“Du”long

beep

Temperature in chamber >150

℃;

The sensor of steam generator does not work.

Check temperature sensor in chamber

2

E32

“Du”long

beep

Temperature of Chamber heater>280

℃;

The sensor of steam generator does not work.

Check temperature sensor of the chamber heater

3

E51

“Du”long

beep

Temperature in chamber 

0

℃;

The sensor of steam generator does not work.The room
temperature is too low.

Check temperature sensor in chamber
Check room temperature.

4

E52

“Du”long

beep

Temperature of chamber heater≤0

℃;

The sensor of steam generator does not work.The room
temperature is too low.

Check temperature sensor in chamber
Check room temperature.

5

E63

“Du”long

beep

1

steam generator temperature≤0

℃;

2

steam

generator temperature>230

℃;

steam generator temperature>230

℃;

Steam temperature

control instability

over 230

℃,

steam generator

temperature sensor damaged.

Check steam generator temperature sensor,
control board,  steam generator

6

E2

“Du”long

beep

The sterilization pressure is 40KPA more than designed
pressure. 

134

:210kpa /121

process:110kpa)

vacuum pump damaged.

Check vacuum pump.  Do a vacuum test

7

E61

“Du”long

beep

134

process

 inner temperature >140

 or 121

process

inner temperature >127

℃;

temperature

control instability.

temperature sensor inside chamber damaged.

Check temperature sensor in chamber.

8

E62

“Du”long

beep

Temperature of Chamber heater>155

℃;

temperature

control instability

control board damaged.

temperature sensor is damaged. Circuit board is
damaged.

Check temperature sensor, chamber heater,
control board, chamber heater

9

E41

“Du”long

beep

In preheat period, after 8mins warm-up, temperature
Chamber heater<100

℃;

chamber heater damaged.

Chamber heater is damaged

check the chamber heater.

10

E42

“Du”long

beep

In preheat period, after 8mins steam generator
temperature <110

℃;

heating rod damaged.

Chamber heater is damaged

Check heating rod

11

E5

“Du”long

beep

When the period of “exhaust”, after working 10mins, the

pressure in chamber still over 0.5bar

air relief

instability.

Outlet solenoid valve is blocked.

Check water drain valve

12

E6

“Du”long

beep

The door is opened during working

the door detector

switch damaged.

Door sensor is damaged

Check door sensor.

13

E7

“Du”long

beep

The local air pressure is  < 70KPa

local air pressure is too low. Wrong air pressure.

Can not use in these area. Adjust atmospheric
pressure:see 5.2

14

E8

“Du”long

beep

In pre-vacuum period, every 5mins temperature raise <
3

.

Air leakage. No water inside tank.

Check water pump, heating rod, control
board.Check water level inside tank.

15

E9

“Du”long

beep

In sterilization period, the sterilization pressure is 0.3 bar
lower than the designed pressure.

No water inside tank. Heater inside steam generator is
damaged. Water pump damaged.

Check heaters, water pump, water level.

16

E10

“Du”long

beep

The electronic locker is in wrong condition

Electronic locked damaged. Circuit board damaged.

check electronic locker and  circuit board

17

E11

“Du”long

beep

The electronic locker is in wrong condition

Electronic locked damaged. Circuit board damaged.

check electronic locker and  circuit board

18

E12

“Du”long

beep

The vacuum not reach -70Kpa 2 times during the
program which have at least 3 times vacuum.

Vacuum pump damaged.Fan damaged.

Check vacuum pump, fan

20

E99

“Du”long

beep

The communication between CPU is wrong.

Connect disabled. CPU loosed.

Check control board data line, and CPU

Summary of Contents for LFSS12AA

Page 1: ...Instruction Manual Autoclave Pressure Steam Sterilizer LFSS12AA LFSS18AA LFSS23AA 1023...

Page 2: ...Revision History Revised Chapter Content Revised Date of Revision Revised By Number of Revision Remarks Please appoint specific person to operate and maintain the device The person must be well traine...

Page 3: ...Introduction 4 2 Using Range 4 3 Parameters and Sterilization Programs 5 4 Operation Panel 6 5 Installation 13 6 Operation 14 7 Abnormal Situations 16 8 Maintenance 17 9 Transportation And Storage 20...

Page 4: ...toclave Notice Please read the manual carefully before operating the autoclave Please operate the autoclave strictly according to this manual Please keep this manual for reference Please contact your...

Page 5: ...ble please contact local distributor or government to recycle it Symbol for MANUFACTURER Symbol for COMPILES WITH MDD93 42 EEC REQUIREMENTS Symbol for DATE OF PRODUCTION Symbol for SERIAL NUMBER Symbo...

Page 6: ...t the wires and do not put heavy things on the wires Do not put the autoclave on an unstable table Do not block the door of autoclave and cover the ventilation holes and radiation holes Do not put any...

Page 7: ...nd vapor to ensure they will not pollute the inner clean pipes 2 Using Range This autoclave can be used for medical purposes e g In general medical practices dentistry facilities for personal hygiene...

Page 8: ...2500VA The highest working pressure 0 21 0 23MPa The highest working temperature 134 137 Life span of the autoclave 5 years The requirements of transportation and storage Environmental temperature ran...

Page 9: ...eam generator to keep the temperature When the Process Vacuum times Sterilizing time Sterilization temperature Sterilization pressure Vacuum persist time Dry time 121 SOLID 1 20min 121 110kPa 3 121 PO...

Page 10: ...is 2 Kpa when the autoclave s door is opened this pressure is the local air pressure 4 1 6 14 09 00 Time User can set it in menu ADV DATE TIME 4 1 7 26 11 2011 Date User can set it in menu ADV DATE T...

Page 11: ...is program s parameters can be adjusted The sterilization temperature the sterilization time dry time and vacuum times B D test is for hollow A loads test B D test and Helix test is the same test prog...

Page 12: ...will print the sterilization records during the working cycle OFF The printer will not print the records during the working cycle 4 3 3 LANGUAGE ENG English PLN Poland 4 3 4 USER DEFINED SET It is use...

Page 13: ...ogram selection page There are six programs can be adjusted Pic1 Pic 1 Pic2 For example you can adjust the parameters of 134 solid program after that the page of Pic2 will be appearing Three parameter...

Page 14: ...the KEY SOUND option press OK to change ON or OFF ON Sound enabled OFF Sound disabled 4 3 9 PREHEAT It is W in LCD screen When the cursor is before the PREHEAT option press OK to change ON or OFF ON I...

Page 15: ...e time arrived pic9 pic10 pic8 pic9 pic10 4 3 11 Dry It have 3 grades I II III If user find that after a cycle loads are wet you can choose II or III The dry s temperature of grade II and III are high...

Page 16: ...nded to install the autoclave in a well ventilated location Don t block the radiator of the sterilizer Keep the table stable horizontal 5 2 Adjust atmospheric pressure Because different countries have...

Page 17: ...inside is enough for a new sterilization cycle You can stop or continue filling water Second time beep It means the clean water tank is full Please stop filling water now 6 2 If you want to use a flas...

Page 18: ...kinds and quantity of the loads When the cycle is finished LCD will show END together with 3 beeps The user can open the door and take out loads inside now 6 3 6 After using you shall turn off the au...

Page 19: ...in chamber 8 E62 Du long beep Temperature of Chamber heater 155 temperature control instability control board damaged temperature sensor is damaged Circuit board is damaged Check temperature sensor ch...

Page 20: ...will tighten the door i e the door will be closer to the chamber Clockwise turning the door will loosen it Detailed steps 1 Push down the lever a little 2 Rotate the door to a specific angle 3 Releas...

Page 21: ...erilize water tanks every week 8 4 To use ethyl alcohol to clean the inner surface of the autoclave every month 8 5 Every 150 cycles we recommend replacing the germ filter 8 6 Replace the fuse 1 Disco...

Page 22: ...is damaged the user must replace it C After the seal ring is cleaned assemble it D Attention if the user finds it hard to put the seal ring back use screwdriver press it carefully until it is done 8...

Page 23: ...torage tank bleeder valve Then turn the drainage switches towards anti clockwise 9 3 Terms of transportation Please refer to sales contract 9 4 Terms of storage The autoclave should be stored in the c...

Page 24: ...The damage is caused by incorrect installation 2 The damage is caused by fall down or impact by carelessness 3 The damage is caused by customer s install or repair 4 Without invoice and guarantee car...

Page 25: ...Cable with plug LF 9 12 2 1 4 Tray shelf see table 11 2 1 5 Tray hand holder LF 9 12 4 1 6 Fuse 6X30 230V20A 2 7 Manual 1 Table 11 1 parts for different models Part Model LFSS12AA LFSS18AA LFSS23AA T...

Page 26: ...23 Appendix 1 Structure diagram...

Page 27: ...24 Appendix 2 Circuit diagram...

Page 28: ...Steam sterilizer uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11...

Page 29: ...s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment interruptions and voltage variations on power supply input line...

Page 30: ...lated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 d 1 2 80 MHz to 800 MHz d 2 3 800 MHz to 2 5 GHz where P is the maximum output power rating...

Page 31: ...help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Steam sterilizer as recommended below according...

Reviews: