9
: Cursor
User can press button “up” or “down” to move
the cursor.
When the cursor is next to the option which user wants to set up
, user can press “OK” to
set up.
4.3.1
、
KEEP TEMP
This is the
“■K” in
LCD screen.
✓
ON: The sterilizer will warm up the chamber and steam generator to keep temperature. When the door is opened,
this function is stopped. The longest time to keep temperature is 8 hours. Setting it on can shorten the time of the
whole cycle.
✓
OFF: Close this function.
4.3.2
、
PRINTER
✓
ON: The printer will print the sterilization records during the working cycle.
✓
OFF: The printer will not print the records during the working cycle.
4.3.3
、
LANGUAGE
✓
ENG: English
✓
PLN: Poland
4.3.4
、
USER-DEFINED SET
It is used to set up the parameters of
“USER
-
DEFINED”
menu. When the cursor is moved to this option, user can enter it
by pressing “OK”. The detailed setting page
is as below:
When the cursor is moved to
this option, User can press “UP” or “DOWN” to change the setting.
Press “OK”
, the cursor will move to the next option.
VACUUM TIMES:
It is used to set up the vacuum times during the sterilization cycle.
✓
04: Meaning that: the program “USER
-
DEFINED” have 4 times vacuum process.
✓
The range of the “VACUUM TIMES” is 1~10.