Lafayette Instrument Flicker Fusion 12021A User Instructions Download Page 4

3700 Sagamore Parkway North  .  Lafayette, IN 47904 USA  .  Phone: (765) 423-1505

Lafayette Instrument Flicker Fusion

Operating Instructions

1. 

With all power off, connect the DB15 cable from the CONTROL connector on the rear of the viewing 

chamber to the VIEWING CHAMBER connector on the rear of the control chassis.

2. 

Connect the black hand-held subject response swtich to the 

external response connector on the rear of the viewing chamber.

3. 

If external stimulus initiation is desired, connect the silver hand-

held push button to the EXTERNAL INITIATE connector on the 

rear of the control chassis.

4. 

If computer control is desired, connect the serial cable from 

the COMM.PORT connector on the rear of the control chassis to 

communication ports 1 or 2 of the computer.

5. 

Once  the  above  connections  are  completed,  plug  the  wall 

mount AC adapter plug into the DC POWER jack on the rear of 

the control chassis.

6. 

Familiarize yourself with the following menu driven controls

7. 

In each mode the subject may respond by pressing either side 

of the black subject response key.

A.  Luminance Adjust (Figure 1)

This adjustment is only available upon powering up of the device.  The stimulus luminance may be adjusted 

using push button #1 to increase and push button #2 to decrease.  Push button #3 will continue to the mode 

selection.  The luminance percentage represents the percent of the maximum stimulus luminance available.  

Even though the stimulus lamps have been pre-aged to minimize decay, as the lamps age some decay in 

luminance will be present, however the left and right stimulus lamps will decay at the same rate.  

Figure 1: Luminance Adjust

B.   Mode Selections (Figure 2)

1.   Ascending Auto Frequency (>AUTOFREQ)

 

 

 

 

 

 

 After selecting luminance or reset, select his mode with push button #1.  Once the mode is selected, select 

the desired stimulus for activation using push buttons #1 to #4 in Figure 3.  Once the stimulus is selected, 

select the desired sweep rate for the test using push buttons #1to #4 in Figure 4.  Once the sweep rate is 

selected, the display will show the mode that has been selected and the lower limit frequency, Figure 5.  

Push buttons #1 to #4 will now perform the functions labeled below the buttons.  To increase the frequency 

setting use #1, to decrease the frequency setting use #2, to initiate the test use #3, and to reset back to 

the mode selection screen use #4.  Once initiated, the selected stimulus will activate and flicker at the 

lower limit.  The flicker frequency will increase at the selected sweep rate until a response or reset occurs.  

When a response is made the frequency is halted and the stimulus is deactivated.  Press #3 to reset for 

the next test.  Press #3 again to reset back to the mode selection window.  If the upper limit frequency of 

100.0 Hz is reached or the reset button is pushed, the stimulus is deactivated and the frequency is reset 

to the starting lower limit.

4

Hand-held Subject 

Response Switch

Summary of Contents for Flicker Fusion 12021A

Page 1: ...nstructions Flicker Fusion User Instructions Model 12021A MAN083 3700 Sagamore Parkway North Lafayette IN 47904 USA Tel 765 423 1505 800 428 7545 Fax 765 423 4111 info lafayetteinstrument com www lafa...

Page 2: ...elf Control Method are available when connected to a computer running Lafayette Instrument Company s PsymLab software available from Lafayette Instrument Company A serial to USB converter is required...

Page 3: ...er variables reported to affect results include body temperature practice general physiological state age sex and genetic variables Specifications Power Supply 10 VDC 1 5A 2 1mm center positive DC plu...

Page 4: ...the percent of the maximum stimulus luminance available Even though the stimulus lamps have been pre aged to minimize decay as the lamps age some decay in luminance will be present however the left a...

Page 5: ...labeled below the buttons To increase the frequency setting use 1 to decrease the frequency setting use 2 and to reset back to the mode selection screen use 4 The stimulus is on continuously in this m...

Page 6: ...perate Window C Stimulus Selection 1 Left Only left stimulus flickers 2 Right Only right stimulus flickers 3 Coincident Both stimuli flicker at the same time 4 Alternate Left and Right stimulus flicke...

Page 7: ...ponse of flicker The half moon shape with solid lines extending out symbolizes a response of fusion In Self Control Method PsymLab uses the Continuous Mode to display a starting frequency and allows t...

Page 8: ...warranty repairs Damaged Goods Damaged instrumentation should not be returned to Lafayette Instrument prior to a thorough inspection If a shipment arrives damaged note damage on delivery bill and have...

Reviews: