Model 12021A User Instructions

Overview

The Model 12021A Flicker Fusion is designed to test for the critical flicker fusion threshold (CFF).  CFF can be used 

as an index of the temporal resolving power of the human visual system.  The threshold of the flicker fusion is 

determined in the following manner: a light beam is interrupted intermittently by electronic means at a slow 

rate, causing it to flash or flicker.  If the rate of the flicker exceeds a certain rate, the light will appear steady.  The 

rate at which flicker ceases and the light appears steady is the threshold of fusion.  The rate of flicker is then 

decreased from steady to the point at which the light changes from steady to flicker, the threshold flicker.  The 

average of the threshold of fusion and the threshold of flicker is called the critical flicker fusion threshold (CFF).

CFF is sensitive to a number of presentation and observer variables.  Presentation variables include stimulus 

frequency, luminance of the stimulus, stimulus size, light/dark ration, stimulus color, and contrast.  Observer 

variables reported to affect results include body temperature, practice, general physiological state, age, sex, 

and genetic variables.

Specifications

Power Supply:

   10 VDC @ 1.5A, 2.1mm center positive DC plug

Fuse:

    0.5A 5 x 20mm fast blow

Frequency:

  1.0 to 100.0 Hz in 0.1 Hz increments with an error of 0.05%

Slide Holder: 

 2” x 2” 35mm holder for optional 12100 neutral density filters with 0.1% to 50% light transmission

Auto Mode Ramp Rates:

 options of 0.5, 1, 2, and 4 Hz per second

Analog Input:

 3.5mm mono phone plug with voltage range from 0.1 to 10V for 1.0 to 100.0 Hz flicker rate; 

ABSOLUTE MAXIMUM INPUT = 14V

External Initiate:

 SPST normally open hand-held switch with RCA input

External Response: 

SPST normally open hand-held switch with 3.5mm stereo plug

RS-232C Port Settings: 

9600 baud, no parity, 8 data bits, 1 stop bit

Typical Maximum Luminance: 

58 Cd/m

2

 

Viewing Angle: 

1.9

o

 

Light/Dark Ratio: 

1:1

Stimulus Color: 

White

Viewing Chamber Mask: 

hypo-allergenic black silicone; mask may be cleaned with an alcohol wipe

Control Size: 

8 5/8”W x 6.5”L x 3 ¼”H; Weight 1.8 lbs

Viewing Chamber Size: 

7”W x 19”L x 16”H; Weight 7.4 lbs

Parts Included

• 

1 – Control Unit

• 

1 – Viewing Chamber

• 

1 – 10 VDC wall mount AC adapter

• 

1 – DB9 Female to Female cable

• 

1 – DB15 Male to Female cable

• 

1 – Single hand-held push button

• 

1 – Dual hand-held push button

3

Summary of Contents for Flicker Fusion 12021A

Page 1: ...nstructions Flicker Fusion User Instructions Model 12021A MAN083 3700 Sagamore Parkway North Lafayette IN 47904 USA Tel 765 423 1505 800 428 7545 Fax 765 423 4111 info lafayetteinstrument com www lafa...

Page 2: ...elf Control Method are available when connected to a computer running Lafayette Instrument Company s PsymLab software available from Lafayette Instrument Company A serial to USB converter is required...

Page 3: ...er variables reported to affect results include body temperature practice general physiological state age sex and genetic variables Specifications Power Supply 10 VDC 1 5A 2 1mm center positive DC plu...

Page 4: ...the percent of the maximum stimulus luminance available Even though the stimulus lamps have been pre aged to minimize decay as the lamps age some decay in luminance will be present however the left a...

Page 5: ...labeled below the buttons To increase the frequency setting use 1 to decrease the frequency setting use 2 and to reset back to the mode selection screen use 4 The stimulus is on continuously in this m...

Page 6: ...perate Window C Stimulus Selection 1 Left Only left stimulus flickers 2 Right Only right stimulus flickers 3 Coincident Both stimuli flicker at the same time 4 Alternate Left and Right stimulus flicke...

Page 7: ...ponse of flicker The half moon shape with solid lines extending out symbolizes a response of fusion In Self Control Method PsymLab uses the Continuous Mode to display a starting frequency and allows t...

Page 8: ...warranty repairs Damaged Goods Damaged instrumentation should not be returned to Lafayette Instrument prior to a thorough inspection If a shipment arrives damaged note damage on delivery bill and have...

Reviews: