Troubleshooting for Leakages
If the device test has failed, check whether the system is leaking. Run the Device Test at
each step, until you find the failure.
Note
If the LSU does not pass one or more of the steps in this test after suggested actions are
taken, the device might need to be returned for service (see the Troubleshooting guide).
Test the
Pump System
Run the Device
Test whilst blocking
the outlet. If the
device passes the
test, there are no
leakages in the
Pump System.
Test the
Vacuum Tube
Run the Device
Test whilst
blocking the
Vacuum Tube. If
the device passes
the test, there are
no leakages in the
Tube.
Test the Patient
Tubing Inlet
Run the Device
Test whilst blocking
the Patient Tubing
inlet on the
Canister. If the
device passes the
test, there are no
leakages in the
Canister.
Device Test
Run the Test