7 Troubleshooting

English

12

Troubleshooting

Warning

Potential for Electrical shock. Do not attempt to open or disassemble pump or electrical accessories.

LCSU 4 will not power “ON”

Verify that the Battery is properly installed and charged.

Check alternative 12V DC or AC sources.

LCSU 4 runs, no Vacuum, or insufficient Vacuum

Verify proper Canister and Tubing connections.

Check system for leaks in Canister and Tubing connections.

Check Canister overflow protection float (800 ml).

Check if  the Filter has been shut-off  (300 ml).

Vacuum too high or too low.

Turn Vacuum Regulator to increase or reduce vacuum level.

Battery will not charge

Verify that the Battery is installed and connected.

Re-connect source power and observe charge (all connections).

Battery seems insufficient

Charge the Battery for 5 hours. 

Perform the Battery Test.

Important

If  the LCSU 4 condition is not resolved, contact Laerdal Medical or one of  its authorized distributors for advice.

Summary of Contents for 880051

Page 1: ...sch Bedienungsanleitung Italiano Istruzioni d uso Nederlands Gebruiksaanwijzing Português Instruções de utilização Svenska Bruksanvisning Suomi Käyttöohjeet Dansk Brugsvejledning Norsk Bruksanvisning 3ROVNL VND µZNL GRW F ÇFH stosowania ˤ ˍˡˤˠ ˢ ˮ ˤ ѣᮽ ֵ 日本語 使用説明書 穢剳檺 斲殯昪律昢 Český Návod k použití Íslenska Notkunarleiðbeiningar LCSU 4 ...

Page 2: ......

Page 3: ...cy Filter Kit 6 Assemble 300 mlVersion 7 4 User Instructions Check Before Each Use 7 Power Source Options 8 Control Panel and Indicator Symbols 8 Suction Level Setting 8 How to Operate and Adjust Suction level 8 5 Battery Information Battery Charging 9 Battery Test 9 Charging Operations 10 External Battery Charger 10 6 Cleaning and Maintenance Cleaning 11 Device Test 11 7 Troubleshooting 12 8 Main...

Page 4: ... DC Adapter charger w AC Power cord Battery Directions for Use Carry Bag for 800 ml version Wire Stand Vacuum Tube LCSU 4 300 ml Cat No 880061 Items Included LCSU 4 Main Unit 300 ml Disposable Canister Patient Tube 0 9 m 3 AC DC Adapter charger w AC Power cord Battery Directions for Use Carry Bag for 300 ml version The main suction unit is the same for both versions Each version can easily be conv...

Page 5: ...table for use in the presence of flammable liquids or gases danger of explosion or fire Do not use the LCSU 4 under environmental conditions that are outside ranges specified This can endanger safety and adversely affect device operation Ingress of suctioned material into the pump can damage and or disable the device If suctioning of liquid from the Canister or patient into the pump is suspected d...

Page 6: ...00 ml Canister has an internal Filter in the Lid The Canister is disposable and cannot be cleaned The Filter automatically stops suction flow when the Canister is full or the Filter becomes saturated if the unit tips onto its side during use Caution Always use the 800 ml Canister supplied by Laerdal which has an internal Filter Never connect any type of Patient Tubing directly to the LCSU 4 Vacuum...

Page 7: ...pt to install or use LCSU 3 Canisters on LCSU 4 Check Before Each Use 1 The Unit should not be damaged 2 The Unit should be clean 3 All parts should be properly assembled Canister Tubes etc 4 Perform Device Test after each reassembly see Chapter 6 5 Check Battery level While performing the Device Test the Battery level should not illuminate RED If the Battery level illuminates RED the Battery shou...

Page 8: ...he AC DC Adapter Charger becomes warm Control Panel and Indicator Symbols Suction Level Setting Green LED Yellow LED Red LED ON OFF switch Suction Level Setting The Scale illuminates Green to indicate the level of vacuum suction strength The Light Blue area indicates reduced suction levels for infants and small children The LEDs have two brightness levels Half illuminated indicates a halfway vacuu...

Page 9: ...rform Device Test see Chapter 6 5 Place Battery on charge see Chapter 5 Battery Charging Important Use only Laerdal Battery Cat No 886113 Caution Use of other than Laerdal brand Battery may result in errors related to the Battery status indicator reduction of the Battery operation time failure to effectively operate LCSU 4 and or give rise to hazards to operator and or patient When in operational ...

Page 10: ...witches off Battery is fully charged Green LED illuminated Connected to AC DC External Battery Charger Cat No 886112 The Battery can be charged externally by removing it from the LCSU 4 unit and using the External Battery Charger Power cord Battery Connector Battery External Battery Charger LED Indicator An empty Battery must charge up to 5 hours to reach full capacity Follow the LED Indicator and...

Page 11: ...e Carry Bags Wipe Bags according to instructions provided above for Main Cabinet Do not launder Device Test After each reassembly and before returning a LCSU 4 to operational use the Device Test should be performed 1 Start the test with a fully charged Battery 2 Turn the unit ON 3 Block the Patient Tube 4 Set Suction Vacuum level to 550 mmHg 5 Unblock and then Block the Patient Tube again 6 Result...

Page 12: ...system for leaks in Canister and Tubing connections 3 Check Canister overflow protection float 800 ml 4 Check if the Filter has been shut off 300 ml Vacuum too high or too low Turn Vacuum Regulator to increase or reduce vacuum level Battery will not charge 1 Verify that the Battery is installed and connected 2 Re connect source power and observe charge all connections Battery seems insufficient 1 ...

Page 13: ...Quantity 886108 Wire Stand for 800 ml Qty 1 886115 Wire Stand for 300 ml Qty 1 88006005 Carry Bag for 300 ml Qty 1 884600 Carry Bag for 800 ml Qty 1 886111 AC DC Adapter Charger w o Power cord Qty 1 886112 External Battery Charger Qty 1 884500 DC Power cord Qty 1 886107 Replacement Filter Qty 10 886116 High Efficiency Filter Kit Qty 1 886108 886115 886111 Spare Parts Cat No Item Quantity 886113 Ba...

Page 14: ... with care IP12 The degree of protection provided by the chassis according to IP12 p Keep dry h Unique product type identification l 70 C 158 F 40 C 40 F Transport storage temperature range f Serial number Humidity 4NG8 UL Classified Atmospheric pressure Center positive polarity indicator Direct Current N Date of production Warning Caution Important Recycle Type BF applied part Disposal must be ca...

Page 15: ...Indicator accuracy 5 of full scale Filter Specifications Internal Filter of the 300 ml Disposable Canister Porous plastic filter PE material Internal Filter of the 800 ml Disposable Canister Aerostate HEPA rated bacterial filter White unfinished surface cellulose with wet strength resin High Efficiency Filter Kit With the High Efficiency Filter Kit installed the unit is in accordance with ISO 1007...

Page 16: ...ss IIa Electrically powered medical suction equipment for field and transport use according to ISO10079 1 1999 High flow High vacuum 50 550 mmHg Not suitable for the use in the presence of flammable liquids or gases Internally powered class I equipment type BF according to IEC 60601 1 Protection class IPI2 and standard power supply Intermittent Operation 30 minutes on 30 minutes off Certifications...