7 Troubleshooting
English
12
Troubleshooting
Warning
Potential for Electrical shock. Do not attempt to open or disassemble pump or electrical accessories.
LCSU 4 will not power “ON”
1
Verify that the Battery is properly installed and charged.
2
Check alternative 12V DC or AC sources.
LCSU 4 runs, no Vacuum, or insufficient Vacuum
1
Verify proper Canister and Tubing connections.
2
Check system for leaks in Canister and Tubing connections.
3
Check Canister overflow protection float (800 ml).
4
Check if the Filter has been shut-off (300 ml).
Vacuum too high or too low.
Turn Vacuum Regulator to increase or reduce vacuum level.
Battery will not charge
1
Verify that the Battery is installed and connected.
2
Re-connect source power and observe charge (all connections).
Battery seems insufficient
1
Charge the Battery for 5 hours.
2
Perform the Battery Test.
Important
If the LCSU 4 condition is not resolved, contact Laerdal Medical or one of its authorized distributors for advice.