Overview of LED/Laser Classifications according to EN 60825-1 and IEC 60825-1 

The IEC (International Electrotechnical Commission) has devised an international standard (IEC-
60825 -1) for the classification, requirements and use of lasers. 
 

Class 3R 

This class is not regarded as dangerous (low risk). The limit for class 3R is 5 mW. 
 
The Laeq Health has a power capacity of 5 mW and belongs to the class 3R lasers. To guarantee a 
power capacity of 5 mW, a power control unit has been built-in that prevents the device from 
ever crossing that 5 mW boundary.  
 
A specific diverging lens is also installed in our laser clip, equally spreading the monochrome 
laser beam.  
 
The optimal amount of energy that is absorbed by the mitochondria is 2 to 4 Joules per cm

2

. This 

can be calculated with the formula J/cm

2

 = W x T, which is published in numerous scientific 

studies. If the intensity is higher than the 2 to 4 Joules per cm

2

 mark, the laser treatment could 

have a negative effect. 
 
To guarantee an optimal amount of energy, our laser is controlled with an app that is 
programmed to never cross that 4 Joules per cm

2

 boundary. The laser cannot be activated when 

the maximum amount of time per day has been reached. 
 
The placement of the clip in the nose and the corresponding maximum treatment time has been 
empirically set. During a period of 30 minutes, every cell in the blood will be radiated 12 times. 
 
The use of the Laeq Health is always safe because: 

 

The 5 mW boundary cannot be crossed – power control 

 

Overdosing is impossible – Laeq Health App 

 

It cannot cause addictions – Laeq Health App 

 

Misuse is impossible – Laeq Health App 

 

10

 

Environment and preventing unauthorized use 

Laeq Health strives to dispose of the used components in a 100% environmentally-friendly 
manner. We also want to prevent incorrect use of the components. We therefore ask you kindly 
to return any components that have been used after the 6 month period is over. You can use the 
included return envelope for this purpose, at no additional costs.  

11

 

Troubleshooting 

Problem 

Possible cause 

Solution 

The Laeq connector does not 
turn on 

The Laeq connector battery is 
empty 

Fully charge the battery 

Summary of Contents for Health 3.0

Page 1: ...Laeq Health 3 0 Product manual Please read the product manual in its entirety to ensure safe use Healthier every day Energized every day Fitter every day Better every day...

Page 2: ...rands or suppliers Only use the Laeq Health for its intended purpose Thank you very much for choosing the Laeq Health This product is a completely safe method that only leads to optimization of your b...

Page 3: ...ndicator 10 QR code 11 Return envelope 12 Laeq carrier 13 Silica gel cartridge 14 Mobile device with Android or iOS operating system not included 15 Laeq Health app free to download from Google Play o...

Page 4: ...e used by multiple people 6 Do not use the Laeq Health while charging the Laeq connector 4 Application of the Laeq Health 4 1 First use 1 Prior to first use fully charge the Laeq connector To do this...

Page 5: ...ector 4 and turn the Laeq connector on with the on off switch 7 The blue Bluetooth indicator 8 and red battery indicator 9 will blink alternately 3 Check whether or not Bluetooth has been activated on...

Page 6: ...aeq connector has been established when the blue Bluetooth connector 8 blinks 7 Place the Laeq PBM clip on your nose pushing it in as far as you can while still being comfortable see picture 8 Start t...

Page 7: ...stop the session the session is unfinished which will lead to diminished results IMPORTANT do not shut down the app during the session keep the smartphone in close proximity to the Laeq connector 10 A...

Page 8: ...e device Consult with your practitioner or contact the Laeq Health customer service 4 3 Using multiple Laeq Health subscriptions on the same smartphone or tablet You can use multiple Laeq Health subsc...

Page 9: ...ponents from other sources can damage the device This also invalidates any warranty agreements 3 Do not open or repair any of the components in case of malfunctioning or for any other reason This will...

Page 10: ...ntee an optimal amount of energy our laser is controlled with an app that is programmed to never cross that 4 Joules per cm 2 boundary The laser cannot be activated when the maximum amount of time per...

Page 11: ...ad the product manual The product is of type BF BF stands for Body Floating This means that any power electricity can never flow into the user Environment protection This device may not be disposed of...

Page 12: ...applied does not automatically lead to an increase of the warranty period for the device or the replaced components 4 The following cases are excluded from the warranty a All damage incurred from inc...

Reviews: