Lactate Pro 2 Operating Manual Download Page 3

- 6 -

7.1 Storage

Store your meter, test strips, lancing device, lancets and manuals in your 
carrying case.

Do not

 freeze. Avoid heat, humidity and direct sunlight.

Storage temperatures:
• Meter: 0 to 50 °C (32 to 122°F)
• Test strips: 1 to 30 °C (34 to 86°F)

7.2 Cleaning Your Meter

This product does not need special cleaning.
If your meter gets dirty, wipe it with a soft piece of cloth moistened with water.
To disinfect the meter after cleaning, wipe with a soft cloth moistened 
with 70% ethanol, 70% isopropanol or diluted household bleach (0.05% 
sodium hypochlorite solution).

7.3 Changing the Battery

When the battery symbol appears on the screen, the battery is getting low.
Before using your meter, change the battery. Past results remain 
in the memory even when the battery is changed.

 

Your meter uses one CR2032 3V lithium battery. This type of battery is 
available in many stores. Keep a spare battery handy at all times.

You do not need to set the date and time if you insert a new battery 
within 2 minutes after removing the old battery. Even if you change the 
battery within 2 minutes, your meter's clock may be reset if:
• You do not replace the battery soon after the battery symbol appears.
• You insert the battery upside down.
• You touch the metal parts inside the meter with hands or metal.

When the time has been reset during battery replacement, "12:00" or 
"0:00" will flash on the screen the next time you insert a Test Strip. In 
this case, set the correct date and time (see chapter 6 "Meter Setup").

1.

 Make sure the meter is off.

2.

 Open the battery cover by lifting the tab with your finger or fingernail.

  Do not touch the metal parts inside the meter with your hands or metal.

3.

 Remove the old battery. Slide the new battery into the battery cover 

with the “+” side facing downward.

• If the battery is inserted upside down, the meter will not operate. 

4.

 Match the lines on the battery cover and the meter. Close the battery cover.

Dispose of old batteries according to your local environmental 
regulations.

Chapter 7  Maintenance

NOTE

NOTE

Error

message

What's the

problem

What to do

Symbol

What's the

problem

What to do

If a problem occurs with the meter or test strips, the beeper sounds and 
the following error codes or symbols appear on the screen.

Chapter 8  If a Message Appears

Contact your distributor.

Leave the meter and test strips in a 
place where the temperature is 5 to 
40°C (41 to 104°F) and humidity is 20 to 
80% RH for at least 20 minutes.
Test your blood only after the error code 
disappears. If the error remains on the 
screen, contact your distributor.

Data could not be sent or received 
during communication. Try again. If the 
error still appears, contact your 
distributor.

Contact your distributor.

Repeat the test with a new test strip.

Use a Lactate Pro

TM

 2 Test Strip to 

repeat the test.

Use a new test strip to repeat the test 
with the correct method.

There is a problem 
with the inside of the 
meter.

The surrounding 
temperature is too 
low or too high.

Communication error

There is a problem 
with the inside of the 
meter.

• The insertion 

portion (electrode) 
of the Test Strip is 
dirty.

• A used Test Strip is 

inserted into the 
meter.

A different type of 
test strip is inserted 
into the meter.

• The test strip 

touched blood 
again after the test 
started.

• The test strip inside 

the meter moved 
during testing.

E-1

E-2

E-3

E-5

E-6

E-7

Change the battery. See chapter 7.3.

Leave the meter and test strips in a 
place where the temperature is 5 to 
40°C (41 to 104°F) and humidity is 20 to 
80% RH for at least 20 minutes.
Test your blood only after the 
thermometer symbol disappears.
Your meter can test your blood even 
when the thermometer symbol is 
displayed, but this may produce 
inaccurate test results.

The battery is getting 
low.

The surrounding 
temperature is too 
low or too high.

- 3 -

1.

 Make sure that the  flashing blood symbol and the date/time appear 

on the screen (see chapter 2.7 "Inserting Test Strips into the Meter").

2.

 Touch the tip of the test strip at a 90 degree angle to the drop of 

blood. 
Let the test strip draw up blood until the check window is filled with 
blood.

• 

Do not

 apply blood when a full display appears on the screen.

• 

Do not

 add any extra blood even immediately after the first draw. It 

may cause inaccurate test results.

Accurate test results will not be obtained if you apply your blood to 
the check window directly.

• For accurate test results, touch the test strip to your blood immediately 

after puncturing.

• 

Do not 

test blood that runs or spreads out from the puncture site.

• 

Do not 

smear blood on the test strip.

• 

Do not 

press the test strip into your puncture site with force. 

• Keep your hands away from the test strip until the test result appears.

3.

 You will hear a beep when the countdown starts from 15 to 1.

4.

 Read your test result.

When "Hi" or "Lo" appears as the test result, it means that the result is 
outside the test range. 
• "Hi": The blood lactate level is above 25.0 mmol/L (225 mg/dL).
• "Lo": The blood lactate level is below 0.5 mmol/L (5 mg/dL).

When the test result is not what you expect, check that:

• The thermometer symbol is not on,
• The meter and the test strip adjusted to the specified temperature 

and humidity before the test,

• The test strip was not exposed for a long time after being removed 

from the aluminum pack,

• The test strip is not expired,
• The test strip is not reused,
• Sweat is not mixed with blood,
• The test strip is filled with enough blood.

5.

 Pull the test strip straight out of the meter.

The meter stores the test result into the 
memory and turns itself off.

• Even if the test strip is not pulled out, the meter will store the test result 

in the memory after 3 minutes and turns itself off.

• New results beyond the 330th result overwrite previously stored results 

in the order of the oldest first, regardless of the measurer code.

Disposal of Infectious Waste

The test strips, lancets, and alcohol cotton pads qualify as infectious 
waste once used to test the blood lactate level. Dispose of them 
according to your local regulations on biohazardous waste.

5.1 Measurer Code

A measurer code makes it possible to display or delete the test results 
by measurer and to manage the data efficiently.

You can add a measurer code before or after test.

 Adding a measurer code before test

1.

 Insert the test strip into the meter. 

Make sure that the blood symbol, date and time appear.

2.

 Press button 1.

The measurer code "M01" appears.

To select a different measurer code, press button 1 again. 
The measurer code changes in the following order every time you 
press the button:

 

3.

 Select a measurer code and test your blood. After the test, the test 

date and the measurer code appear alternately. 

4.

 Pull the test strip straight out of the meter.

 Adding a measurer code after test

1.

 When the test result appears, press button 1 

without removing the test strip.
The measurer code "M01" appears.

To set a different measurer code, press button 1 again.
The measurer code changes in the following order every time you 
press the button:

2.

 To save the setting and turn off the meter, pull the test strip straight 

out of the meter.

• You cannot add/change the measurer code once the meter turns off.
• The meter turns off when you remove the test strip or after 3 minutes 

from testing.

Chapter 5  Managing Your Test Results

Chapter 4  Testing Your Blood Lactate

NOTE

NOTE

NOTE

NOTE

M01

M02

M03

(no m. code)

M01

M02

M03

(no m. code)

NOTE

Enough

Not enough

Reviews: