2.7 Inserting Test Strips into the Meter
Insert the test strip into the meter as described below.
1.
With clean, dry hands, remove 1 test strip from
the aluminum pack.
•
Do not
cover the check window with your fingers.
•
Do not
apply blood to the test strip before inserting it into the meter.
2.
Insert the test strip fully into the test strip slot.
You will hear a beep when the meter turns on. A full display appears.
A flashing blood symbol and the most recent test result appear on the
screen.
Start blood sampling when the current date
and time appears.
• For the mg/dL spec meter, "mg/dL" appears instead of "mmol/L".
• A test result with a thermometer symbol or a delete flag does not
appear as a previous test result.
• When 4 minutes and 30 seconds passes after inserting a test strip,
you will hear a beep every 5 seconds. The meter turns off
automatically after another 30 seconds. In this case, remove the test
strip once and insert it back into the test strip slot.
• If nothing appears on the screen, remove the test strip and insert it
back into the test strip slot. If there is no change, replace the battery. If
this does not work, contact your distributor.
• When an error code appears, refer to chapter 8 "If a Message
Appears" and take the necessary steps.
2.6 Lactate Pro
TM
2 Test Strip
NOTE
3.1 Cautions on Blood Sampling
3.2 Blood Sampling Method
For accurate test results, perform blood sampling as described below.
1.
Disinfect the puncture site with alcohol cotton
pads, etc.
2.
Let the alcohol dry thoroughly.
Make sure the puncture site is dry. Otherwise it may lead to hemolysis.
3.
Puncture with the lancing device.
The method for using a lancing device is different for each one. Please
read the manual that comes with the device.
4.
Press lightly around the puncture site to obtain
a small drop of blood.
5.
Use clean gauze, etc. to wipe off the blood once.
6.
Press lightly around the puncture site again to obtain a small drop of
blood.
Chapter 3 Blood Sampling
• Use extreme care when handling blood. Incorrect handling can
cause you or people around you to be infected by pathogenic
microorganisms.
Follow these warnings to reduce the risk of infection.
WARNING
•
Do not
share the same lancet or lancing device with anyone to
avoid the risk of infection by pathogenic microorganisms.
• Always use a new lancet. Lancets are for single-use only.
Do not
reuse a lancet that you have already used.
• When puncturing a thin part of the body such as an earlobe, etc.,
do not
directly support the rear of such part with your finger. (There
is a risk of puncturing your finger with the needle penetrating the
part of the body. This may cause an infection via blood. When there
is a possibility of penetrating through to the other side, consider
puncturing a thicker part of the body.)
CAUTION
NOTE
NOTE
NOTE
Always disinfect the puncture site to avoid the risk of infection.
After sampling blood, stop bleeding and protect the puncture site.
WARNING
Check window
- 7 -
*
1
The meter is shipped with the battery loaded. So, the battery may not
satisfy the "Battery life".
*
2
The data can be stored on a personal computer using software for this
meter. Please contact your distributor for details.
Product specifications and appearance are subject to change for
improvements without notice.
Chapter 9 Technical Information
Lactate Pro
TM
2
LT-1730
Blood lactate level
Fresh capillary whole blood
0.3
μ
L
Lactate Pro
TM
2 Test Strip
0.5 to 25.0 mmol/L (5 to 225 mg/dL)
15 seconds from blood detection
3 V lithium battery (CR2032) × 1
2000 tests or more
*
1
(Actual number of tests may be less under some
conditions of use.)
0.02 W (MAX.)
330 test results
(A maximum of 330 test results for all three
measurer codes combined.)
Yes (volume setting possible in 4 levels)
Yes
*
2
19200 bps
Temperature: 5 to 40°C (41 to 104°F)
Humidity: 20 to 80% RH (No condensation)
0 to 50°C (32 to 122°F)
3 years (according to company data)
50 mm (W) × 100 mm (H) × 12 mm (D)
Product
Model
Test item
Sample
Sample size
Test strips
Test range
Test time
Battery
Battery life
Power
consumption
Memory capacity
Beeper
Communication
function
Communication
speed
Operating
environment
Storage
temperature
Expected life
Dimensions
9.1 Specifications
9.4 Symbols
9.5 Warranty
ARKRAY shall repair or replace the Lactate Pro
TM
2 free-of-charge in the
event that there are defects in material and workmanship for a period of
one year from the date of purchase. However, this warranty does not
apply to the following:
1.
Trouble or damage due to careless use.
2.
Trouble or damage due to unforeseen circumstances such as natural
disasters.
3.
Trouble or damage due to unauthorized repairs or remodeling.
4.
Trouble or damage for which ARKRAY is not responsible.
9.2 Product Safety Information
Electromagnetic Interference (EMI)
This meter complies with CISPR 11: 2003, Class B (Radiated Only). It
emits low levels of energy that are not likely to interfere with nearby
electronic equipment.
Static Electricity and Radiated Magnetic Field Immunity
This meter clears immunity requirements for Level 3 electrostatic
discharge set forth in IEC 61000-4-2. This meter clears immunity
requirements for radio frequency interference in the 80 MHz to 2.5 GHz
range (3 V/m) set forth in IEC 61000-4-3.
9.3 Disposing of Your Meter
The meter qualifies as biohazardous waste once used to test blood
lactate. When no longer needed, remove the battery and dispose of the
meter according to your local regulations on biohazardous waste. This
meter is not subject to requirements of European Directive 2002/96/EC
(Directive on Waste Electrical and Electronic Equipment (WEEE)).
Storage temperature
limitation
In Vitro Diagnostic
Medical Device
Manufacturer
Biological risks
Caution, consult
accompanying
documents
Batch code
Use by
Symbol
Description
Lactate Pro
TM
2
conforms to the Directive
98/79/EC
Suitable for self testing
Authorized
Representative in the
European Community
Serial number
Consult instructions for
use
Catalogue number
Symbol
Description
Issued: 2012.11
0
50
Summary of Contents for Pro 2
Page 1: ......