5. Installation
●
Caution during installation
- Do not install in an inclined or uneven ground.
- Install the equipment horizontally and approximately 15cm away from the wall.
- Establish minimum spacing required for inspection and repairs.
- Do not leave flammable or volatile materials near the equipment such as gasoline.
- Do not put anything on top of the equipment.
- Ambient temperature 5℃ to 40℃, Optimum temperature 18℃ to 25℃
- Allowance of supplying voltage change: lower than 10%
* Wrong power supply (over/less voltage) may cause serious damage to compressor and freezer.
The damage from wrong power supply is not covered under warranty.
If the power supply is unstable, please install AVR (Automatic Voltage Regulator) separately.
- For electric safety, the unit should be earthed.
If the power socket is fitted with earth cord, please check the outlet for earth before use.
* Do not earth to a gas pipe, a telephone line or lightning conductor.
●
When not used for a long time.
- Unplug the power cord from the outlet when the equipment is not used for long hours and to
keep it in safe from moisture.
●
When transporting
- Do not lay the equipment on its sides when moving
If the equipment is moved to a long distant location, it may cause a malfunction.
●
Do not dismantle/remodel
- Do not dismantle, repair, or remodel unless you are an authorized repair technician.
There is danger of fire, electric shock, and injury due to abnormal operation.
●
Transport & storage condition
- Temperature: 0℃ to 70℃
- Humidity: 0% to 90%
- Pressure: 700hPa to 1060hPa