4.2. Power and safety device

 

 

 
 
 
 
 
 

  Over Temp 

Over temperature protection device 
- It is a dial type temperature controller. It automatically cuts the heater output off when there is 
over temperature due to equipment malfunction. 
- Set this 10~20% higher than the set temperature. 
- Over temperature protection device is analog type; therefore, it will not accurately match the 
displaying temperature. 

②  Power switch 

Power on/off switch

 

③  Circuit Breaker 

Device to supply power and to cut off an overcurrent 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

① 

② 

③ 

Summary of Contents for LBI-150E

Page 1: ...1 Operating Manual Number LBI 500EE Edited December 1999 Revised January 2016 July 2017 May 2018 Low Temperature BOD Incubator Model LBI 150E LBI 250E LBI 500E LBI 1000E...

Page 2: ...described in this manual and operation procedures are not taken properly it may result in malfunction of the equipment Note This mark is to indicate precautions which must be taken properly for safe...

Page 3: ...match it may cause fire or electric shock and if used in combination with other equipment it may cause abnormal heat generation and fire The figures diagrams and numeric values used in this manual are...

Page 4: ...of contents 1 Tables of contents 4 2 Product overview 5 3 Product specifications 6 4 Part and functions 7 5 Installation 9 6 Operating 10 7 Maintenance and repair of products 12 8 Cause and trouble s...

Page 5: ...bility ensure user safety This equipment equipped with a compressor which allows general incubation as well as low temperature incubation experiments This equipment is made with high quality stainless...

Page 6: ...0 1 Uniformity 0 5 at 20 Controller Microprocessor PID Multi Function Controller Display 4 Digits LED Display Timer 99hr 59 min Continuous Material Inner Stainless Steel SUS304 Outer Epoxy Powder coat...

Page 7: ...is in operation Heater operation indicating lamp OUTPUT Illuminates the heater is in operation Timer display lamp TIMER Illuminates when timer is operating START STOP button START STOP Button to start...

Page 8: ...heater output off when there is over temperature due to equipment malfunction Set this 10 20 higher than the set temperature Over temperature protection device is analog type therefore it will not acc...

Page 9: ...lease install AVR Automatic Voltage Regulator separately For electric safety the unit should be earthed If the power socket is fitted with earth cord please check the outlet for earth before use Do no...

Page 10: ...erature will be indicated on the display waiting for input 4 Press LEFT SHIFT button to move to digit you want change value 5 Press INC button to change value 6 4 How to set time 1 Press MODE button a...

Page 11: ...ication lamp illuminates and the timer starts to count down 3 After set time Timer indicating lamp turns off and operation stops 6 7 Auto tuning 1 Auto tuning function automatically sets optimal P ID...

Page 12: ...the compressor If the sound of equipment operating suddenly becomes loud or irregular please check with our service team 7 2 Maintenance of electric compartment Be careful not to let water enter the e...

Page 13: ...eaker short circuit check Switch on the circuit breaker When temperature control is not possible Check the condition of the controller Calibrating regulator settings Over temperature protection check...

Page 14: ...14 APPENDIX Consumables List NO PART CODE NAME SPECIFICATION UNIT Q ty 1 LBI E001 HEATER LBI 150E M 750W EA 1 LBI 250E M 1KW EA 1 LBI 500E M 750W EA 2 LBI 1000E M 750W EA 3 2 3 4 5 6 7 8 9 10...