6. component description

 

 

 

Test stand housing (A)

 

Thanks to its ergonomic design, the test stand 

can be operated fatigue

-

free even during con-

tinuous operation. Operation in computer mo-

de is via PC keyboard and mouse or manually 

via the optionally available front panel. 

 

 

Front panel (B) (optional)

 

The front panel contains the control buttons 

and a multi

-

function LCD display showing print 

and system information.

 

The following abbreviation is used:

 

LP = Low pressure (

-

50...+50 mbar)

 

MP = medium pressure (0...25 bar)

 

 

Multifunction probe (C)

 

The test head is a fixed head that can be infla-

ted by an electric pump so that full face masks 

of any make can be sealed without problems. 

A measuring point is installed in the test head 

which measures the internal pressure in the 

mask or lung automat.

 

 

Medium pressure device connections (D, E)

 

Plug nipple for connecting the medium pres-

sure line of the compressed air breathing ap-

paratus pressure regulator to be tested. 

 

Quick coupling for connecting the medium 

pressure line of the regulator to be tested.

 

 

CPS test connection and system vent (F)

 

Quick coupling for filling CPS. For all other test 

procedures, the coupling must be connected 

to the supplied plug

-

in nipple/silencer for sys-

tem venting.

 

 

Interface (H)

 

The USB interface for data transfer to a PC is 

located on the left side of the housing 

(LabModul 240 MT has no function with the 

manual device).

 

 

Vacuum pump 

 

With the built

-

in vacuum pump, the required air 

currents are generated and the probe is infla-

ted or deflated. The desired air flow rate can 

be set in the software from 2...12 l/min.

 

 

Processor 

 

The built

-

in microprocessor takes over the 

control and signal conversion. 

 

 

 

 

 

 

 

 

7. description of the operating elements

 

 

 
 
 
 
 
 

 

 

1

 

2

 

 

 

LCD multi function display

 

Digital display of status of operation

 

Display for low pressure (LP)

 

Display for medium pressure (MP)

 

3

 

Push button for test head inflation

 

Pressing causes the test head to be inflated out 

of the test head

 

4

 

Push button for

 

test head aspiration 

 

Pressing causes the air to be aspirated out of 

the test head

 

5

 

Push button to generate positive pressure

 

Forcing pump +5

 

l/min is set in motion by press-

ing this button

 

6

 

Push button to generate negative pressure

 

Forcing pump –

5

 

l/min is set in motion by press-

ing this button

 

7

 

Push button for low pressure decompress-

ing

 

Pressing causes the air to be aspirated out of 

the test equipment

 

8

 

Switch button for forcing pump

 

Forcing pump +10

 

l/min is set in motion by 

pressing this button. LED shows function. 

Pressing button again switches it off.

 

 

9

 

Switch button for forcing pump

 

Forcing pump 

-

10

 

l/min is set in motion by 

pressing this button. LED shows function. 

Pressing button again switches it off.

 

10

 

Switch button for CPS inflation

 

Pressing causes the CPS to be inflated out of 

the CPS, Medium pressure is needed in the test 

bench.

 

11

 

Switch button for CPS deflation

 

Pressing causes the air to be aspirated out of 

the CPS

 

12

 

Push button for medium pressure decom-

pression

 

Pressing causes the air to be aspirated out of 

the test equipment

 

Summary of Contents for LabModul 240 Com

Page 1: ...2 3 device description 3 4 technical data 3 5 operating elements 4 6 component description 5 7 description of the operating elements 5 8 commissioning 6 9 care and maintenance 6 Attachments Performin...

Page 2: ...ed and certified Labtec service technician and tested and calibrated by him or her using a Labtec calibration device The list of current certified employees training partners and service technicians i...

Page 3: ...fitted at the factory Connection requirements A 230 240 V 50 60 Hz power supply is requi red to operate the test stand As medium pressure connection Euro coupling and nipple are available as standard...

Page 4: ...10 l min switching valve 9 Inhalation 10 l min switching valve 10 Fill CPS switching valve 11 Empty CPS switching valve 12 Relieve medium pressure tactile valve A test stand housing B Panel optional C...

Page 5: ...um pump With the built in vacuum pump the required air currents are generated and the probe is infla ted or deflated The desired air flow rate can be set in the software from 2 12 l min Processor The...

Page 6: ...g nipple of the medium pressure device connection and the line of the demand valve to the EURO quick coupling of the medium pres sure device connection If the SCBA to be tested does not have a Euro co...

Page 7: ...specific adaptors 1 2 2 Mount mask on test head pull straps tight inflate test head 3 A few seconds is usually long enough 1 2 3 Adjust negative pressure to exactly mbar 5 By means of pressure relief...

Page 8: ...with the setpoint All further tests require high pressure Regulator in conjunction with a Connect the SCBA to the test device Medium pressure hose from the PA regulator of a SCBA must with Device conn...

Page 9: ...ssure flows into the device 3 2 2 Close the bottle of the SCBA Tester DUT are under pressure 3 2 3 Read the medium pressure on the LCD display MP and compare with setpoint 3 2 4 After a specified time...

Page 10: ...pressure is reached 5 1 8 Lower the pressure to the required internal pres sure with the switch 11 if necessary with but ton 12 approach and turn on the stopwatch Setting the required Settling time 5...

Page 11: ...h in EEC Guideline 2016 42 EG and EMC 2014 30 EU including all changes and addendums to date thereto The above mentioned product meets the following standards and guidelines Nous Vous Confirmons que l...

Reviews: