User Manual LabModul 240

 

Page 10

 

www.labtec.de

 

5

 

Leakage

 

5.1

 

Leakage of the suit

 

 

6

 

A

-

Valve

 

6.1 

Tightness test of suit valves No. 1 to 4 after # min. at 

-

 # mbar  

 

6.4

 

(Repeat test according to the number of valves )

 

Test 

step

 

Operation

 

Comments

 

Controls

 

5.1.1

 

The suit

-

tightness test requires 

 

medium pressure !

 

Connect the CPS to the tester in conjunction with 

a SCBA. 

 

Medium pressure hose from the regu-

lator of a SCBA must with

 

Device connection (D) to be connec-

ted. 

 

D

 

5.1.2

 

Connect a tightening valve with test adapter, 

seal all other tightening valves with plugs. 

 

CPS test adapter CGA 200 and manu-

facturer

-

specific adapter 

 

 

5.1.3

 

Engage filling hose of the test adapter with plug 

nipple in filling connection (F). 

 

CPS test adapter CGA 200 

 

F

 

5.1.4

 

Engage filling hose of the test adapter with plug 

nipple in filling connection (C).

 

CPS

-

test adapter CGA 200

 

C

 

5.1.5

 

Open the bottle of the SCBA .

 

Medium pressure flows into the device.

 

 

5.1.6

 

Press the switch (10) to inflate CPS. 

 

CPS is filled to> # mbar .

 

 

5.1.7

 

Press the switch (10) again to stop filling the suit. 

 

when the required internal pressure is 

reached .

 

 

5.1.8

 

Lower the pressure to the required internal pres-

sure with the switch (11), if necessary with but-

ton (12) approach and turn on the stopwatch. 

 

Setting the required Settling time. 

 

 

5.1.9

 

If time has expired, lower switch (11), if necessa-

ry with button (12) to the required test pressure 

and switch on stopwatch. 

 

 

 

5.1.10

 

After a specified time of # min read off the low 

pressure on the LCD display (LP) and compare 

with the previously set test pressure. 

 

Compare difference value with allo-

wable deviation. 

 

 

5.1.11

 

Close the bottle of the compressed air respirator 

and relieve the test system with the push button 

(12), remove the test adapter 

 

Open suit.

 

 

 

 

 

 

6.1.1

 

6.4.1

 

Engage filling hose of the test adapter into the 

mouth opening of the test head (C). 

 

CPS test adapter  CGA 200

 

C

 

6.1.2

 

6.4.2

 

Connect suit valves 1 to 4 one after the other 

with valve test adapter from the inside 

 

brand

-

specific adapter,

 

depending on the number of the suit

 

existing valves . 

 

 

6.1.3

 

6.4.3

 

Test vacuum with pushbutton (6) to setpoint

 

-

 Set # mbar. Turn on the stopwatch. 

 

If necessary, relieve with button (7) 

and approach to setpoint. 

 

 

6.1.5

 

6.4.5

 

After a specified time of # min. Read pressure 

change on the LCD display (LP) and compare 

with setpoint. 

 

 

 

A 4    Testing gas

-

tight chemical protective suits  (CPS)

 

Summary of Contents for LabModul 240 Com

Page 1: ...2 3 device description 3 4 technical data 3 5 operating elements 4 6 component description 5 7 description of the operating elements 5 8 commissioning 6 9 care and maintenance 6 Attachments Performin...

Page 2: ...ed and certified Labtec service technician and tested and calibrated by him or her using a Labtec calibration device The list of current certified employees training partners and service technicians i...

Page 3: ...fitted at the factory Connection requirements A 230 240 V 50 60 Hz power supply is requi red to operate the test stand As medium pressure connection Euro coupling and nipple are available as standard...

Page 4: ...10 l min switching valve 9 Inhalation 10 l min switching valve 10 Fill CPS switching valve 11 Empty CPS switching valve 12 Relieve medium pressure tactile valve A test stand housing B Panel optional C...

Page 5: ...um pump With the built in vacuum pump the required air currents are generated and the probe is infla ted or deflated The desired air flow rate can be set in the software from 2 12 l min Processor The...

Page 6: ...g nipple of the medium pressure device connection and the line of the demand valve to the EURO quick coupling of the medium pres sure device connection If the SCBA to be tested does not have a Euro co...

Page 7: ...specific adaptors 1 2 2 Mount mask on test head pull straps tight inflate test head 3 A few seconds is usually long enough 1 2 3 Adjust negative pressure to exactly mbar 5 By means of pressure relief...

Page 8: ...with the setpoint All further tests require high pressure Regulator in conjunction with a Connect the SCBA to the test device Medium pressure hose from the PA regulator of a SCBA must with Device conn...

Page 9: ...ssure flows into the device 3 2 2 Close the bottle of the SCBA Tester DUT are under pressure 3 2 3 Read the medium pressure on the LCD display MP and compare with setpoint 3 2 4 After a specified time...

Page 10: ...pressure is reached 5 1 8 Lower the pressure to the required internal pres sure with the switch 11 if necessary with but ton 12 approach and turn on the stopwatch Setting the required Settling time 5...

Page 11: ...h in EEC Guideline 2016 42 EG and EMC 2014 30 EU including all changes and addendums to date thereto The above mentioned product meets the following standards and guidelines Nous Vous Confirmons que l...

Reviews: