6. operating elements
LCD multi function display
Digital display of status of operation
Display for low pressure (LP)
Display for medium pressure (MP)
Display for high pressure (HP)
Push button for test head inflation
Pressing causes the test head to be inflated out of the test head
Push button for
test head deflation
Pressing causes the air to be aspirated out of the test head
Push button to generate positive pressure
Forcing pump +5
l/min is set in motion by pressing this button
Push button to generate negative pressure
Forcing pump –
5
l/min is set in motion by pressing this button
Push button for low pressure relief
Pressing causes the air to be aspirated out of the test equipment
Switch button for 10 l/m exhalation
Forcing pump +10
l/min is set in motion by pressing this button. LED shows function. Pressing
button again switches it off.
Switch button for 10 l/m inhalation
Forcing pump
-
10
l/min is set in motion by pressing this button. LED shows function. Pressing
button again switches it off.
Switch button for CPS inflation
Pressing causes the CPS to be inflated.
Switch button for CPS deflation
Pressing causes the CPS to be deflated.
Push button for medium pressure relief
Pressing causes the air to be aspirated out of the test equipment
Switch button for the main valve
When pressed, the high pressure main valve is opened which raises the high pressure. Pressing
button again switches it off.
Push button for high pressure relief
Pressing causes the air to be aspirated out of the test equipment
Push button for Medium pressure increase (Only active with option SI)
When pressing the medium pressure is increased until the safety valve on the device opens.
Switch button for Artifical Lung (only active with ECO dyanmic, main valve has to be open)
When pressed, the artificial lung is switched on and ventilated through the mouth opening of the
test head. Pressing button again switches it off.