INSTALLATION AND OPERATING INSTRUCTIONS
1.
Installation
a.
Receiving Inspection:
Upon receipt, check all packages and accessories or optional components including legs,
casters, and shelves.
b.
Legs & Casters: (See Fig. 1)
Legs screw directly into the base. Casters are screwed directly into the base.
c.
Leveling:
To provide adequate condensate drainage and proper door alignment and operation. It
is necessary that the cabinet be level. This should be done after cabinet has been set in
its final operating position. BE SURE THE DRAIN HOSE IS ATTACHED TO BRACKET NEXT
TO EXTERIOR CONDENSATE PAN LOCATED BEHIND BOTTOM GRILLE. THE HOSE COULD
BE DISLODGED IN TRANSIT.
d.
Shelves:
Shipped inside each cabinet are shelves packed in plastic and a bag of shelf supports.
Shelf spacing is adjustable to suit requirements. See Instruction Sheet ILA-0904 for
installing.
e.
Door Removal and Adjustment:
(1) Slide Doors – Each door has its own Closing Spring located at the top and is
positioned for Proper Tension. However if adjustment is required remove doors as
follows:
a.
Locate Door Rollers under slot A show in fig.4, and lift door off track while tilting
bottom of door out.
Adjust spring by moving hook to next hole or by shortening
If door gaskets do not seal properly follow step 1 and 2
Step 1: Check to see if cabinet base is level.
Sept 2: Adjust the Location of door Rollers in the Roller Brackets.
(2) Swing Doors – Spring tension in the bottom hinges is Factory Adjusted for adequate
“Swing Back” of doors, and does not require field adjustment. For replacement parts
and instructions see enclosed instruction sheet ILA-0859.
f.
Locating Cooler:
Provide at least three inches of space between cabinet and any adjacent wall or fixture,
at left end and rear of cabinet.