1.

 

Installation 

a.

 

Receiving Inspection: 
Upon receipt, check all packages and accessories or optional components including legs, 
casters, and shelves. 
 

b.

 

Legs & Casters: (See Fig. 1) 
Legs screw directly into the base. Casters are screwed directly into the base. 
 

c.

 

Leveling:  
To provide adequate condensate drainage and proper door alignment and operation. It 
is necessary that the cabinet be level. This should be done after cabinet has been set in 
its final operating position. BE SURE THE DRAIN HOSE IS ATTACHED TO BRACKET NEXT 
TO EXTERIOR CONDENSATE PAN LOCATED BEHIND BOTTOM GRILLE. THE HOSE COULD 
BE DISLODGED IN TRANSIT.  
 

d.

 

Shelves: 
Shipped inside each cabinet are shelves packed in plastic and a bag of shelf supports. 
Shelf spacing is adjustable to suit requirements. See Instruction Sheet ILA-0904 for 
installing.  
 

e.

 

Door Removal and Adjustment: 
(1) Slide Doors – Each door has its own Closing Spring located at the top and is 
positioned for Proper Tension. However if adjustment is required remove doors as 
follows: 
a. 

Locate Door Rollers under slot A show in fig.4, and lift door off track while tilting 

 

bottom of door out.  

Adjust spring by moving hook to next hole or by shortening 
If door gaskets do not seal properly follow step 1 and 2 
Step 1:  Check to see if cabinet base is level. 
Sept 2:  Adjust the Location of door Rollers in the Roller Brackets.  
(2) Swing Doors – Spring tension in the bottom hinges is Factory Adjusted for adequate 
“Swing Back” of doors, and does not require field adjustment. For replacement parts 
and instructions see enclosed instruction sheet ILA-0859.  
 

f.

 

Locating Cooler: 
Provide at least three inches of space between cabinet and any adjacent wall or fixture, 
at left end and rear of cabinet.  

 

Summary of Contents for LABL-27-HG

Page 1: ...n includes complete refrigeration System cabinet construction finish IMPORTANT PLEASE RETAIN FOR YOUR RECORDS LABL 27 HG LABL 38 SG LABL 45 SG LABL 49 HG AND LABL 74 HG 101 Witmer Road Suite 700 Horsh...

Page 2: ...ustable to suit requirements See Instruction Sheet ILA 0904 for installing e Door Removal and Adjustment 1 Slide Doors Each door has its own Closing Spring located at the top and is positioned for Pro...

Page 3: ...r flow over the pan hastens condensate evaporation so that external drain plumbing is not required 3 Maintenance Caution Disconnect Electrical service before accessing mechanical enclosures having mov...

Page 4: ......

Page 5: ......

Page 6: ...LL SCALE SHELF SUPPORT FRONT WITH 1 4 TAB FULL SCALE NOTES INSTRUCTIONS 1 PLACE SHELF SUPPORTS WITH 1 4 TABS INTO PILASTER AT DESIRED POSITION NOTE THAT SUPPORTS WITH 1 4 TABS ARE LOCATED AT FRONT ONL...

Page 7: ...hields by using the identification mark on the end of the bulb Properly installed bulbs will have the mark centered between the halves of the lamp holder On units with shields the identification mark...

Page 8: ......

Page 9: ...sy Off Oven Cleaner Apply generous coating Excellent removal fatty acids miikstone Allow to stand for 10 to 15 Satisfactory for use on all where swabbing or rubbing minutes Rinse Repeated finishes is...

Page 10: ...if necessary Check air flow across condenser Unit runs OK but short cycles 1 Overload protector 1 Check wiring diag for correct wiring 2 Cold control 2 Differential set too close 3 Overcharge 3 Reduce...

Reviews: