Sensor Errors
In addition to any alarms, the control will take the following action if it detects any of the
unit’s sensors as being in error
Alarm
Result
Control Probe failure
Both compressors will turn on and remain on.
Heat exchanger probe failure
Compressor control during pull down will revert to timed
staging
Condenser probe failure
Alarm only, no change in compressor control
Ambient probe failure
Alarm only, no change in compressor control
Sample probe failure
Alarm only, no change in compressor control
Line/boost voltage sensing
Unit will operate normally, with boost voltage circuit Off