Luna Wireless Module
© 2008
– 2020 by MMS B.V.
119
8.3.3
Set the investigation protocol
In the protocol settings of the stationary measurement program you will find
the channel configuration by clicking on the
Connections
button. The channel
names will pre-fixed by the Luna module.
The default settings are:
Pves (1)
= Luna P1
Pabd (2)
= Luna P2
Pura (3)
= Luna P3
EMG1
= Luna EMG
Conductance1
= Luna Conductance
8.3.4
Connect the catheter
Solid state or air charged transducers with Lemo 4 plug can be connected to
the Lemo 4 inputs for measuring pressures. In case you have a catheter with
conductance rings, connect the Lemo 3 plug to the conductance input on the
Luna. The color of the air charged sensors represent the pressures as follows:
Color air charged catheter
Pressure
Yellow
Pves
Blue
Pabd
Green
Pura
The pressure sensors of the air charged catheter must be left in the
‘open’
position.
Summary of Contents for Solar Blue
Page 1: ......
Page 9: ...8 2008 2020 by MMS B V...
Page 19: ...18 2008 2020 by MMS B V Chapter 1...
Page 47: ...46 2008 2020 by MMS B V Chapter 3...
Page 75: ...74 2008 2020 by MMS B V Chapter 4...
Page 81: ...80 2008 2020 by MMS B V Chapter 5...
Page 138: ...2008 2020 by MMS B V 137 APPENDICI WARRANTY AND INDEX...
Page 139: ...138 2008 2020 by MMS B V...
Page 157: ...156 2008 2020 by MMS B V Appendix A...
Page 165: ...164 2008 2020 by MMS B V Appendix C...
Page 173: ...172 2008 2020 by MMS B V Appendix D...
Page 174: ...2008 2020 by MMS B V 173 Appendix E Solar Blue Safety Barriers...
Page 175: ...174 2008 2020 by MMS B V Appendix E...
Page 189: ...2008 2020 by MMS B V Warranty for Solar Blue System...
Page 198: ......