4
LABORIE Portascan3D Service Manual PTS3D-SM01
V3.0
3. UNIT QUALITY INSPECTION
3.1 ACCURACY
A phantom is used to test the accuracy of volume readings.
Inside the phantom is a simulated bladder with a volume
mentioned on each unit.
Take 5 scans and check if the highest and lowest one from the
5 readings are within the tolerance.
Caution! Please tune the gain value up to 40dB when being
inspected on the phantom, and tune the gain value back down
to 20dB after the inspection!
NO.
REQUIREMENTS
REMARKS
1
Tolerance
≤±15%
40 dB Gain on Phantom
20 dB Gain on Human
3.2 FUNCTIONS
Test each function in routine operations according to the user’s
manual.
In this test item, we focuses on touch screen button, probe
button, interface display and printing.
NO.
REQUIREMENTS
REMARKS
2
Touch Screen Button
Probe Button
Interface Display
Printing
3.3 APPEARANCE
This test item is focused on color, text labeling, scratch marks
and spots.
Check whether plastic parts bubble, crack, deform or overflow,
whether operational and adjustment mechanisms are flexible
and secure, and whether fastening pieces are tightly fastened.