SL 20i
GUIDANCE TABLES (continued)
The SL 20i is intended for use in the electromagnetic environment specified below. The
customer or user of the SL 20i should ensure that it is used in such an environment.
21
Table 202 – Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
- Guidance
ESD
IEC 61000-4-2
±6kv Contact
±8kv Air
±6kv Contact
±8kv Air
EFT
IEC 61000-4-4
±2kv Mains
±1kv I/Os
±2kv Mains
±1kv I/Os
Surge
IEC 61000-4-5
±1kv Differential
±2kv Common
±1kv Differential
±2kv Common
voltage
Dips/Dropout
IEC 61000-4-11
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
3A/m
3A/m
8126700-795
SL 20i
Issue 1.0
Printed on June,2019
Floor should be wood, concrete or ceramic tile. If
floors are synthetic, the R/H should be at least 30%.
Mains power quality should be that of a typical
Commercial or hospital environment.
Mains power quality should be that of a typical
Commercial or hospital environment.
Power frequency magnetic fields should be that
of a typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the SL 20i requires continued
operation during power mains interruptions,
it is recommended that the SL 20i be powered
from an uninterpretable power supply or
battery.